FDA Device Recalls in 2018

773 recalls found

December 2018

FDA

MOSAIQ Oncology Information System Version 2.64. MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery

If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for imaging when it is sent to the machine.

Elekta notified customers on about 12/04/2018 via letter titled, "Urgent Important Field Safety Notification." The letter contains recommended user action and that appropriate personnel working with the device should be advised on the content of the letter and the letter should be placed in an area accessible to all users. The letter stated that the issue will be resolved in Service Packs to MOSAIQ Releases 2.64, 2.65 and 2.70. Customers will be informed when the fix is available through a Product Bulletin. An acknowledgement Form was sent as part of the letter to be completed and returned no later than 30 days. The firm can be contacted at 408-830-8000 or 855-693-5358.

FDA

Alaris System PC Unit Model 8015 with software versions 9.17 and 9.19; Central programming, monitoring and power supply component for the Alaris System.

A patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110.

An Urgent Medical Device Recall Letter was sent to customers on 5/31/16 to inform them that BD had identified an issue with the Alaris System PC unit model 8015 with software versions 9.17 and 9.19. Firm has received reports that a patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110. The letter informs the customers of the potential risk and the required action for users. Customers are informed that BD will contact them within 60 days to initiate the scheduling process for the software upgrade installation.

FDA

GEM Coupler Forceps, GEM4183C, packaged individually.

Potential presence of rust on Coupler Forceps (GEM4183C).

Synovis Micro Companies Alliance (MCS), Inc. a subsidiary of Baxter International Inc., notified customers on about 11/02/2018 via "Urgent Product Recall letter sent via U.S.P.S., first class mail. Instructions included to review inventory for affected product, arrange for return of all affected product to Synovis MCA, complete and return the provided Baxter customer reply form, and notify customers if the affected product was further distributed. General questions can be directed to Synovia MCA at 800-510-3318 or 205-941-0111 Monday through Friday, between 8:00 am and 5:00 pm Central Time.

FDA

Medtronic MiniMed 630G 6025805-003_C, Rx Only Medtronic MiniMed 670G 6025805-005_B, Rx Only Medtronic MiniMed 640G (International)

Reports of occurrences in which insulin pumps with version 4.10 software have failed to make expected audio sounds during alerts, alarms or sirens. This could cause a user to miss system notifications, alarms or sirens associated with how the pump is working, and with high and low glucose alerts.

On 10/3/2018, the firm sent an URGENT FIELD SAFETY NOTIFICATION to customers and distributors via email and/or US Postal Service First Class to inform them of reports of occurrences in which insulin pumps with version 4.10 software has failed to make expected audio sounds during alerts, alarms, or sirens. This failure could either cause the alarm volume to be stuck at a 4 (out of 5) level regardless of the personal setting, or it could make no sound. Either of these occurrences could cause the user to miss system notifications, alarms or sirens associated with how the pump is working, and with high and low glucose alerts. Consignees are being informed that they are receiving this letter because the firm's records indicate they may have one of the potentially affected insulin pumps. The notification further explains, this issue occurs when a specific electronic component in the pump malfunctions. This potential loss of audio could delay the user's response to the underlying reason for the alert, alarm, or siren, which could then lead to possible health and safety risks such as hypoglycemia or hyperglycemia. If this issue occurs with the pump, the audio cannot be permanently repaired or regained. The only way to permanently regain the audio capabilities is to exchange the affected pump with a replacement. Even if the pump has the audio issue, it will continue to deliver insulin as expected. Loss of the Vibrate feature has not been reported to occur as part of the above audio failure mode. If a customer has a pump that contains the specific electronic component, and the pump passes the Audio Beep test described below, there is still a chance that the pump may lose its audio capabilities at some future date. Customers are instructed to: What you should do: 1. If you have not already done so, enable the Vibrate feature on your pump. Even if the Audio Beep test passes, you should enable the Vibrate feature (in addition to your Audio feature) since it will add

FDA

Widex BABY hearing aid (referred to as BABY440), powered by zinc-air batteries size 10 Intended use: BABY440 is intended for, but not limited to, babies (0 to 3 years of age) to be used as air conduction amplification devices in everyday listening environments.

The reason for this field action is due to the mounting of a non-tamper-resistant (ordinary) battery drawer on some BABY440 hearing aids. The correct battery drawer is tamper-resistant and require a special tool to open it.

The firm, Widex, sent an "Urgent Medical Device Correction Letter" dated 8/28/18 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: a) Investigate, whether your childs BABY440 hearing aid is mounted with the correct battery drawer. Please refer to the attachment How to determine if correct battery drawer is mounted. b) If affirmative, return the product to the hearing care professional, where you bought the hearing aid, to have the tamper-resistant battery drawer mounted. If you have any questions, contact Director, Global Regulatory Affairs at 0045 44 35 57 71 or email to: krpe@widex.com.

FDA

Mindray BeneVision Distributed Monitoring System (DMS), Part No. 115-034413-00

The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around between multiple central stations in the same care area.

The firm notified affected consignees on August 30, 2018, via Urgent Medical Device Recall letter. The letter advised customers of the software issue when more than 64 telemetry transmitters have been connected to a central station. Customers were advised to use the system normally, but informed that a Mindray Service Representative would be in contact to arrange for a system software update to be performed at the customer site. Mindrays Technical Support team may be reached at (877) 913-9663 (Option 1), Monday through Friday, 8:30am - 5:30pm ET should you have technical questions. Customers were also asked to complete and return an attached Acknowledgement and Receipt Form.

FDA

BECKMAN COULTER MicroScan MICroSTREP Plus Panel, MicroScan Mueller-Hinton Broth with 3% lysed horse blood, Catalog Number 81015 MicroScan Mueller Hinton Broth with 3% Lysed Horse Blood is used for inoculation of MicroScan MICroSTREP plus and MicroFAST panels. MICroSTREP plus panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae.

Beckman Coulter has received customer complaints of lowered minimum inhibitory concentration (MIC) for antimicrobial agents Cefaclor (Ctr) and Penicillin (P) with control organism Streptococcus pneumoniae (ATCC 49619) during quality control (QC) testing on MICroSTREP plus(R) Panel Type 1 hydrated with Mueller-Hinton Broth with 3% lysed horse blood part number B1015-25 lot number 1005986140.

Beckman Coulter sent an Urgent Medical Device Recall letter dated March 16, 2018. Please discontinue use and discard any remaining inventory of the broth lot listed above. You should retain your inventory of other lot numbers of this broth as they are not impacted by this issue. Follow the guidance in the Broth Preparation section of the panel procedural and QC manual. Beckman Coulter recommends discussing the content of this letter with your medical director regarding the need to review previous test results, conduct patient follow-up, and/or repeat testing by another lot of this broth if the isolates are still available. For further questions, please call (469) 961-4941.

FDA

remel SUPERPACK CS/ 500 SEG/FLASK, REF R01626, Qty:50, plate label: REMEL MH W/OXACILLIN Product Usage: Recommended for the use in qualitative procedures to screen Staphylococcus aureus for resistance to penicillinase- resistance penicillins (e.g., methicillin, nafcillin, and oxacillin).

Product may fail performance testing for S aureus ATCC 43300.

The firm initiated the recall by letter on 05/04/2018. The letter stated the following: "Please notify any personnel who need to be aware of the potential for failure of performance for this product. Accordingly, in keeping with our Quality Policy, we request that you inspect your stock and destroy any remaining inventory of the lots listed above and contact Customer Services or your local distributor regarding replacement. You should review results and laboratory reporting associated with use of the listed lots and consider retesting and/or seek appropriate expert advice for further action." For questions contact Technical Services Department at 800-255-6730 or 913-888-0939.

FDA

Centricity Universal Viewer Zero Footprint Client, Model Number 2090258-001, Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2, SP9 and SP9.0.1 display o orientation markers can become is incorrect.

Issue #1 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a and SP8.0.2. When an image is flipped horizontally and vertically, the orientation markers appear correctly on the image to which they were applied. However, once new image rendering happens on this image or series (with scrolling; playing cin; maximizing or changing viewports/layouts; or synchronizing viewports and resizing the browser), the new image orientation is correctly maintained but the display of orientation markers can become incorrect. Issue #2 affects Software Versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.a, SP8.0.2, SP9 and SP9.0.1. When an image in a series is flipped and then rotated in Zero Footprint Viewer (ZFP), the orientation markers appear correctly on the image to which they were applied. However, once you navigate to another image in the series and the image you navigated to has a different orientation than the prior image from which you navigated, the new image orientation is correctly maintained but the display of orientation markers can become incorrect.

The firm issued the medical device correction notices on 11/09/2018 by letter. The letter outlined the problems and provided work-around strategies to be employed until a product modification is available to correct the issues.

FDA

Endo GIA" Radial Reload with Tri-Staple" Technology The Endo GIA radial reloads with Tri-Staple Technology have application in open or minimally invasive general abdominal, gynecologic and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, e.g. low anterior resection. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pan

The device cartridge disengaged during use due to manufacturing error.

Medtronic sent an "URGENT MEDICAL RECALL LETTER" dated September 13, 2017, was issued to customers titled "Covidien Endo GIA Black Radial Reload with Tri-Staple Technology" urging customers to quarantine and return unused product to recalling firm. Questions or concerns can be directed to: feedback.customerservice@Covidien.com. For further questions, please call (203) 492-5000.

FDA

MAC(TM) Two-Lumen Central Venous Access Kit with ARROWg+ard Blue(R) Access Device, Integral Hemostasis Valve, Sharps Safety Features and Maximal Barrier Precautions for use with 7.5 - 8 Fr. Catheters The MAC(TM) Multi-Lumen Central Venous Access Device with ARROWg+ard Blue(R) permits venous access and catheter introduction to the central circulation. It may be inserted into the jugular, subclavian, or femoral veins. The ARROWg+ard(R) technology is intended to help provide protection against cath

Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.

Arrow International sent an Urgent Medical Device Notification Letter dated April 5, 2017, to affected customers to inform them of the issue. The notice provided safety instructions for using the product and instructed customers to return the enclosed acknowledgement form indicating the notice was received, and fax it to 855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Customers with questions were instructed to call 1-866-246-6990. For questions regarding this recall call 610-378-0131.

FDA

Covidien Emprint Long Percutaneous Antenna with Thermosphere Technology, Material CA30L2 intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.

On November 15, 2018 Medtronic issued UPDATED URGENT MEDICAL DEVICE RECALL notices to customers. Customers are advised to take the following actions: 1. Please immediately quarantine and discontinue use of the affected products. 2. Return affected product as indicated. 3. If you have distributed the products, please promptly forward the information from the letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions or concerns, please do not hesitate to contact your Medtronic representative or Technical Support at (800) 448-3644, Option 3, Option 8.

FDA

ClariVein IC infusion catheter, Model Number: 65-018-E4S Product Usage: The ClariVein IC infusion catheter is indicated for infusion of physician-specified agents into the peripheral vasculature.

Certain catheters could have a compromised catheter shaft due to a manufacturing defect which could result in some delamination of the catheter shaft including during introduction of the catheter into a vascular access introducer.

Vascular Insights sent an Urgent Voluntary Medical Device Recall letter to their customers. The letter identified the affected product, problem and the actions to be taken. Customers were instructed to 1. Do not use or distribute any recalled product. 2. Immediately remove all recalled product from your inventory. 3. Segregate recalled product in a secure location and follow the instruction herein to return the recalled product and receive replacement product. 4. Immediately forward a copy of this recall notification to any/all sites to which you may have distributed the affected product and provide directions to follow the instructions contained herein. Customers were asked to complete and return the enclosed Reply Verification Tracking Form.

FDA

PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered

During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.

The firm initiated their recall by email on 09/08/2017. The consignee was warned that the lift straps may wear prematurely and they would be contacted by a representative and the units would be replaced.

FDA

Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009

A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.

The firm initiated their recall by email on 07/28/2017. The firm directed the consignee to repair the units on site and destroy the affected parts.

FDA

UNIVERSAL FLEX2 BREATHING CIRCUIT, Item Numbers: D460-6133Z, D390-6121Z, D390-6121Z, D495-61Z, DF4116-6121Z, D466-61Z, DFP170-6121, DF3115-6121Z, DF375-6121Z, DF3115-6121Z, DF3115-6121Z, D360-8031Z, D465-61Z, JD365-6033Z, DF370-6121Z, DF3110-61Z, DF375-17621Z, DF375-6121Z, DF475-17621Z, DF370-6121Z, DF375-6121Z, DF470-6021Z, D366-618Z, D390-60Z, D390-61Z, DP260-61, DF370-6121Z, D3120-6154Z, F95100, JD365-6033Z, DF370-61Z, DF475-61Z, DF3110-6121Z, DF3110-6121Z, DF375-6121Z, DF475-6121Z, F90755,

Potential for leak test failures.

On 3/15/2017, the firm issued a letter to its sales force describing its intent to replace product for consignees that complained about leaking. The firm also issued a copy of the manifold IFU that includes the need for a 1/4 turn in the Universal Flex2 Breathing Circuit (with bacterial filter).

FDA

TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00

A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.

Certified letters were sent to all affected customers on 07/14/2017, to offer a replacement of the battery charger.

FDA

Alaris Syringe Pump, Model No. 8110. Delivers fluids. Expansion of Affected Units: BOM Housing Assembly Service, Part No. 49000226, and Lower Housing/Carriage Block Assembly, Part No. 148188-100.

Channel Error code is displayed on the PC unit with an audio and visual alarm, and on the syringe module. After the error is cleared on the PCU, the syringe pump is unresponsive to key presses until the next power cycle, or the module is detached and reattached.

An Urgent Medical Device Recall Notification letter was sent on 7/20/15 to customers who purchased the Alaris Syringe Module (Model 8110) to inform them that CareFusion has identified an issue with the Alaris Syringe module manufactured between March 2014 through September 2014. The letter informs customers of the problems identified and the actions to be taken. CareFusion will replace the syringe drive train assembly on the affected units and adjust, if required. CareFusion will contact all affected customers within 60 days to initiate the scheduling process for the remediation. Customers are instructed to contact CareFusion Support Center (888-562-6018, 7am-4pm PT, M-F, supportcenter@carefusion.com) for recall related questions. Customers are instructed to contact Customer Advocacy (888-812-3266, 24hrs/day, 7days/wk, customerfeedback@carefusion.com) for adverse event reports. Customer are instructed to contact Technical Support (888-812-3229, 6am-5pm PT, M-F, dl-us-inf-techsupport@carefusion.com) for technical questions. Customers are instructed to complete and return the enclosed customer response card to expedite the corrective action process. A urgent recall letter for the expansion of the affected units was sent on 10/15/15 to inform customers of the expanded recall. The letter informs customers that the firm has identified an issue with the Alaris Syringe module model 8110 after receiving reports of customers experiencing an channel error on the Alaris Syringe module Model 8110. The letter informs customers that the recall is an expansion of affected units outlined in the CareFusion recall previously issued on 7/20/15. The letter provides the customers with the affected units, issue, potential risk, and actions to be taken. The corrective action for the expanded serial numbers will not change. The affected Alaris Syringe modules will be remediated with a hardware inspection and an adjustment, if required. CareFusion will contact all affected c

FDA

Vado Steerable Sheath 8.8F, REF SS8FMU67, packaged in a lidded tray that is within a pouch which is inserted into a shelf carton, Sterile, Rx. The firm name on the label is Kalila Medical, Campbell, CA.

Torn sheath liner material was detected at the proximal end of the device during internal product testing which may be the result of damage incurred during the manufacturing process.

The recalling firm issued a letter dated 12/14/2017 via FedEx on 12/14/2017.

FDA

Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.

When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.

The firm initiated the recall by letter on 05/24/2018. The letter explained the issue and requested the UPS(s) be disconnected from the system until the system could be repaired.

FDA

MOOG Curlin Infusion Administration Set, Ref 340-4114, Non-DEHP Microbore Tubing with Non-Vented Bag Spike, packaged 20/box, for use with the MOOG Curlin Ambulatory Volumetric Infusion Pump, RX, Sterile. Product Usage: Administration set to be used with the Moog Curlin infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral infusions.

Particulate found on the Curlin spike.

The recalling firm began issuing letters via USPS certified mail with return receipt on 11/10/2017 and concluded on 11/13/2017. Updated letters dated 11/20/2017 were issued via certified mail to distributors and to medical facilities. The distributor letter emphasized their instructions to notify downstream accounts. The medical facility letter updated lot numbers. The letter described the issue, risk to health, and actions the customer was to take. A response form was enclosed to report whether the customer has inventory they are not returning, that they no longer have inventory of the product, or the amount of product being returned for replacement. The form was to be returned via FAX or email.

FDA

STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi English SRK.. General and Plastic Surgery: The Intuitive Surgical EndoWrist Stapler 45, Stapler 45 Reloads and other Stapler Accessories are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device can be used with staple line or tissue butressing material (n

Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.

Intuitive sent an Urgent Medical Device Correction (ISIFA2017-02-C) letter dated March 16, 2017 were sent by FED EX to be received on March 17, 2017. Letters advised users of the reason for the notification, risk to health, affected products with part numbers and a copy of Appendix A - proper use of the Stapler Release Kit (SRK). Customers were instructed to distribute a copy of the letter to all users at their facility, follow all instructions using the Instrument Release Kit (IRK) tool to finish manually unclamping the stapler and return the SRK to Intuitive Surgical through the Returned Material Authorization (RMA) process. Customers were also instructed to log into the da Vinci Online Community Field Action resource (US only) to read or complete any request actions related to this issue. In the case where the da Vinci online resource cannot be used (including outside he US), complete the attached Acknowledgement Form and return it via fax to Intuitive Surgical as instructed on the form. Customers were asked to retain a copy of the letter with their Stapler User Manual and the acknowledgement form for their files. If you need further information or support concerning this notification , please contact Intuitive Surgical Customer Service at the numbers listed below: North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customersupport-servicesupport@intusurg.com Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or ics@intusurg.com South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) For questions regarding this recall call 408-523-2100.

FDA

IntelliVue MX40 Patient Monitor, Model Nos. 865350, 865351, 865352, 867146 - Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals.

A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.

On November 9, 2018, the firm notified customers of the issue via Customer Information Letters. The letter discussed the product problem. Customers are asked to confirm speaker behavior. It is recommended that the clinical user reboot the MX40 prior to entering Monitoring Mode to confirm proper operation of the speaker. Customers are also asked to complete the attached Reply Card and return to Philips as soon as possible. The firm will replace affected devices used at above 4,500 feet (1,372 meters). If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377.

FDA

PROCARE B20 Patient Monitor, V1; and B20 Patient Monitor, V2. The PROCARE Monitor B20 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intrahospital transport.

Patient monitors may restart due to network overload caused by network configuration.

The recalling firm, GE Healthcare, issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 9/11/2018 via trackable mail on 9/11/2018 to its customers. The letters described the product, problem and actions to be taken. The letters informed the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Bx0 and/or B1x5 patient monitor on the network does not resume normal functionality in approximately 90 seconds. The customers were instructed to follow the safety instructions. GE Healthcare will provide revised network installation and user instructions documentation at no cost to you in a future mailing. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

FDA

Covidien Nellcor SpO2 Forehead Sensor Item Code: RS10; The Covidien Nellcor Forehead SpO2Sensor, model RS10, is indicated for use when continuous non-invasive arterial oxygen saturation and pulse rate monitoring are required for patients weighing more than 40 kg.

Labeling error: The label indicates the product is latex free, but the headband component included with the forehead sensor does contain latex.

Medtronic initiated on December 16, 2016. Customers were notified via Federal Express and the letter instructed customers to discontinue use of the product and return all inventory to Medtronic (Covidien) 195 McDermott Road, North Haven, CT 06473 Attn: Field Returns Department. Complete a Recalled Product Return Form (Attachment A, included with letter) and submit it to customer service so a Returned Goods Authorization number (RGA#) can be issued. Questions contact Medtronic representative or customer service at 800-882-5878.

FDA

Invia Motion Negative Pressure Wound Therapy System. Labeled as: a. Endure Negative Pressure Wound Therapy (Model 087.4010); b. 15 Days Negative Pressure Wound Therapy (Model 087.4012); c. 60 Days Negative Pressure Wound Therapy (Model 087.4014); d. Endure Negative Pressure Wound Therapy (Model 087.4020 ); e. 15 Days Negative Pressure Wound Therapy (Model 087.4022); f. 60 Days Negative Pressure Wound Therapy (Model 087.4024). Product Usage: The Invia Motion Negative Pressure Wou

Device may display a battery missing error.

Medela sent an Notification letter dated February 1, 2018 to customers via Certified Mail. The letter identified the affected product, problem and actions to be taken. Instructions included to place all affected items on hold until an insert outlining additional charging instructions were included with the devices. Those instructions included to provide a printed copy of the charging instructions in each device package, review those insert instructions with patients and care-givers, and complete and return the confirmation letter. For questions contact Medela Partner Services at info-healthcare@medela.com or via phone 877-735-1626.

FDA

iBalance(R) UKA, Femoral Cemented, Size 3, Left Medial/Right LATERAL, REF AR 501-UFLC, QTY 1, Rx ONLY, STERILE R, Arthrex, Inc., Naples, FL Indicated for use in uni-compartmental knee arthroplasty as a result of: 1. Moderately disabling joint disease of the knee resulting from painful osteoarthritis or posttraumatic arthritis. 2. Correction of functional deformity. 3. Revision of previous unsuccessful uni-compartmental knee replacement or other procedure. 4. As an alternative to tibial osteotomy

Arthrex part number AR-501-UFLC, size 3, batch 10123297, has the incorrect part number AR-501 UFRA, size 1 printed on the patient label inside the box.

Arthrex sent an Urgent Medical Device Voluntary Recall letter on October 3, 2017. The letter requested the following: "Immediately discontinue use of these devices. In order to receive credit for the returned devices, please follow the steps below: 1) Review this notification and ensure affected personnel are aware of the contents. 2) Immediately identify and return all the indicated product/batch numbers you have in inventory. Call Arthrex Field Logistics Service at 1-866612-7899 for a returned goods authorization (RGA) and product return instructions. Email: FLS@Arthrex.com." Customers with questions were instructed to call Product Surveillance at 866-267-9138 or by email to commplaints@arthrex.com.

November 2018

FDA

Family Wellness First Aid Sterile Gauze Pad 3 in x 3 in, 10 CT. White, bleached, non-woven, rayon/polyester sterile Gauze Pad 3 inch x 3 inch, 12-ply 10 count cardboard box

Potential that gauze pads may not be fully sterilized

ASO, LLC notified customers of the recall on about 10/16/2018 via "URGENT: MEDICAL DEVICE RECALL" letter sent to corporate. Instructions were to immediately stop using/distributing the affected gauze pads, place them in quarantine, and return to ASO LLC. Customers were also instructed to complete and return the response form.

FDA

Dash 3000 Patient Monitoring System is used to monitor physiologic parameter data on adult, pediatric and neonatal patients.

The patient monitors may simultaneously restart as designed if all are connected to the same network and a network overload occurs for a prolonged time.

The recalling firm issued letters dated 8/31/2018 via FedEx on 8/31/2018 informing the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Dash and/or Solar patient monitor on the network does not resume normal functionality in approximately 90 seconds.

FDA

Stratus CS STAT Fluorometric Analyzer-microprocessor-controlled instrument that measures certain analytes in body fluids for in vitro diagnostic use. Assays include ~hCG, CRP, CKMB, cTNI, D-Dimer, Myoglobin, and pBNP. SMN: 10444834, 10453531

Software defect, where either an Above Assay Range or an inaccurate value could potentially be reported, in the Stratus CS Acute Care Diagnostics System regarding TestPak calibration and quality control (QC) status when a TestPak is due to expire within 48 hours.

Siemens Healthcare issued an Urgent Field Safety Notice by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on June 22, 2016 both in the United States and Outside the US for communication with affected customers. This notice informs customers of the issue and provides mitigation instructions. Customers are to contact Siemens Customer Care Center or their local Siemens technical support representative at 800-405-6473. Siemens Healthcare is currently developing a software update to address this issue and will be providing new information as it becomes available

FDA

Spacelabs Healthcare Xhibit Telemetry Receiver, Model 96280, software version 1.1, physiological patient monitor (with arrhythmia detection or alarms) The Spacelabs Healthcare Telemetry Receiver, Model 96280, is intended to provide the Spacelabs Healthcare monitoring system with adult, pediatric and neonatal patient data of patients connected to Spacelabs Healthcare telemetry transmitters. Data includes physiological waveforms and calculations, cardiac arrhythmia and ST data, and patient demogr

Spacelabs has received multiple complaints reporting telemetry beds dropping off the Spacelabs Central Station resulting in a loss of monitoring.

On January , 2018, a customer letter was sent via priority service, return receipt requested, to all U.S. customers. International consignees will be notified on January 24 , 2018, by email. The letter stated: " Please weigh the benefits versus the risks when deciding whether or not to continue to use your Exhibit Telemetry Receiver(s) until updated as set forth below. We recommend that you brief your staff regarding the failure mode listed above." Spacelabs Healthcare will contact you to schedule a convenient time for Spacelabs to update your Exhibit Telemetry Receiver(s) to software version 1.0.2 at no cost." For further questions, please call (800) 522-7025.

FDA

Medi-Vac(TM) Flex Advantage(TM) Suction Canister Liner, Material Numbers: a) REF 6561-910C b) REF 6561-916K c) REF 6561-920C To remove blood and/or body fluids. These fluids may be suctioned during surgical procedures or other applications requiring the removal of built up fluids. The suction canisters are used in operating rooms but are also used outside the surgical field as well as post-operatively within patient suites. The collection containers hold the fluids until they are ready for prop

products may be missing a one-way valve on the fluid patient port.

Cardinal Health sent an Urgent Product Recall letter dated March 28, 2018, Please take the following actions: 1. INSPECT your inventory for the affected product code and lot number (listed on package) 2. SEGREGATE and QUARANTINE all on-hand product 3. RETURN the enclosed acknowledgment form via fax to 847-689-9101 or 614-652- 9648, whether or not you have affected product, indicating the product code, lot and quantity of product you've quarantined 4. CONTACT Customer Service at 800-964-5227 to arrange for return and credit of any affected product . For further questions, please call (800) 292- 9332.

FDA

Edwards Lifesciences Aortic Perfusion Cannula, 24F x 30CM, resterilized.

The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.

Customers were notified via letter on about 03/02/2018. Instructions included to review inventory for affected devices, arrange for their return to SPS Sterilization, and complete and return the Buisness Response Card. A follow up letter was sent to customer on about 03/07/2018 and included a description of the risk and hazard involved. Instructions remained the same.

FDA

PALLAS M/MAXIMIS 5.5 mm Instrument Tap. Part of MAXIMIS Pedicle Screw Spinal Fixation System. Valorem Surgical, Chicago, IL. Product labeled as Pallas M Spinal Fixation System through 2015. Intended for posterior pedicle screw fixation of the non-cervical spine.

Devices are not consistent with Quality System Requirements.

Valorem Surgical sent an Urgent Medical Device Recall letter dated June 21, 2017, informing consignees of the product removal and instructing consignees to return any product to Valorem Surgical LLC to minimize potential risk to patients. Consignees are to fill out the attached Medical Device Recall Return Response form and send it back to the firm by email at global.safety@ValoremSurgical.com or by fax at 312-275-7117, or by mail. For questions pertaining to this recall, please submit an email inquiry to global.safety@ValoremSurgical.com

FDA

Overhead Counterpoise Systems (OCS), OCS CONF,STD CEIL MNT, STD HORZ ARM, Catalog # OCS 115A, Material Number 59880253

The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.

On June 28, 2018, Urgent Medical Device Field Safety Corrective Action letters were distributed to customers. The letters informed the customers of the following: A Bayer Service Representative will be scheduling a service visit to replace the ceiling column assembly of your OCS system(s) at no cost to you.

FDA

Sedecal NOVA FA DR System. for diagnostic radiography.

A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient, user, or third parties.

Initial communication from SEDECAL to all our distributors was performed on 07/18/2018 by email. The letter explained the problem and requested the consignee cease use of the unit if they observe an abnormal displacement in the vertical axis or any suspicious noise during the vertical movement.

FDA

K-Assay(R) Multi-Analyte Calibrator, REF KAI--016C, KAMIYA BIOMEDICAL COMPANY, Seattle, WA, IVD The K-ASSAY(R) Multi-Analyte Calibrator is intended to be used for the calibration of the K-ASSAY(R) Alpha-1 Acid Glycoprotein, Alpha-1 Anti-Trypsin, Complement C3, Complement C4, Haptoglobin, lgA, lgG, lgM, and Transferrin immunoturbidimetric assays. For in vitro diagnostic use

The AlAT calibrator values in the package insert for the Multi-Analyte Calibrator, lot D755, lots N653 and lot N654 were not correct.

Kamiya Biomedical Company sent an Urgent Medical Device Recall/Correction Notice on July 14, 2017 to affected customers, providing them with updated package inserts with corrected A1AT values and/or free replacement kits. Customers were asked to fill in and return the acknowledgement form. Customers with questions were instructed to contact the firm at 206-575-8068.

FDA

LMA(TM) MADgic(TM) Laryngo-Tracheal Mucosal Atomization Devices: (a) MAD600 MADgic, (b) MAD700 MADgic, (c) MAD720 MADgic Intended for the application of topical anesthetics to the oropharynx and upper airway region

Teleflex Medical is recalling the affected product because there may be missing or incomplete information on the package label.

Teleflex sent an Urgent Medical Device Recall Notification letter on July 5, 2017, to all affected customers. The letter directed the following actions: "1. If you have affected stock, immediately discontinue use and quarantine any products with the product codes and lot numbers listed above. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter." For questions regarding this recall call 866-246-6990.

FDA

Ascom Telligence Nurse Call System

There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.

Ascom Partners / Distributors were contacted by e-Mail, USPS Mail and Phone as necessary on 11/02/2018 to determine where the Telligence 5.0 Product is installed.

FDA

Arietta 70 Ultrasound System Intended for use by trained personnel (doctor, sonographer. etc.) for the diagnostic Ultrasound evaluation.

The strength of the arm mounting mechanism was insufficient and had to be redesigned.

An Urgent: Device Correction Notice was sent via overnight carrier on 5/5/17. The Notice identified the affected product and problem. The firm will replace the Monitor Arm to correct the issue at no cost to the customer. If the problem occurs, customers are to call 800-800-4925 for immediate service. Hitachi Service will update the hardware at each site. Customers can contact recalls@hitachihealthcare.com with questions, comments, or requests regarding the Notice.

FDA

Dymaxeon 5.5 Curved Rod Size 5.5 x 110mm (Yellow), Model Number 06.6441B Product Usage: The Dymaxeon Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral/iliac spine (T1 S1/Ileum): degenerative disc disease (defined as discogenic back pain with degeneration of disc confirmed by history and radiog

Packages labeled as containing 110mm rods may contain 100mm rods.

Customers were contacted via phone the week of 3/5/18 and 7/12/18. The customers were informed of the recall and asked any of the affected remained in their possession, if the products have already been implanted, and if any patients experienced any adverse events.

FDA

Bivona Tracheostomy Tube Tracheostomy Tubes

Carton labeling is printed with "Sterile" indicated labeling, however the device is not sold as sterile.

Customers were notified via letter on approximately 04/02/2018. Instructions included to determine if there are any affected Customized Bivona Tracheostomy Tubes in inventory, notify customers if the devices have been further distributed, arrange for the return of affected devices, and complete and return the acknowledgement response form.

FDA

Battery Power Line II Battery Oscillator, DePuy Synthes Power Tools, The Anspach Effort Inc.

Has the incorrect operating directions on the mode switch ring, i.e. its is labeled "FWD, LOCK, REV" instead of "ON, LOCK, ON".

DePuy Synthes sent an Urgent Notice Recall Notification dated November 24, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They were instructed to review their inventory and immediately remove any affected devices, complete the verification section of the letter and return it to the Anspach Effort Inc., and return any affected product to their DePuy Synthes Sales Consultant. Any questions regarding this recall / removal contact the Complain Handling Unit Manager at (561) 494-3673 or their DePuy Synthes Sales Consultant. For further questions regarding this, please call (561) 627-1080.

FDA

Diamedix Is-CMV IgG Test Kit

Product contained an incorrect substrate.

On March 26, 2015 Erba Diagnostics sent a communication to their clients informing them of the recall and sending t hem a replacement sustrate HRP to replaced the wrong one sent on the kit.

FDA

Powdered Surgical Gloves, labeled as: a. Baxter and Ansell brand Powdered Surgical Gloves b. Powdered Surgical Gloves included in ACS surgical convenience kits Health care personnel rely on medical gloves as barriers against transmission of infectious diseases and contaminants when conducting surgery, as well as when conducting more limited interactions with patients.

Powdered Surgical Gloves are an FDA banned substance.

The firm, ACS, notified Customers via phone call and email on about 12/28/2016. The customers were informed about the product, problem and actions to be taken. The customers were instructed to identify and remove all the affected kits for replacement of the recalled gloves; tray insert label must reflect the omission of the powered gloves; and ban, quarantine and return affected product. If you have any questions, contact that Quality Manager at 816-920-5846 or email to: tbakke@amconsys.com.

FDA

Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388

The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

Customers were notified on about 09/20/2018 via an "URGENT - Medical Device Correction" letter. An initial letter was sent to distributors with instructions to provide the Field Safety Notice to all customers the devices were further distributed to. The "URGENT - Medical Device Correction" letter provided guidance on how to identify the Service Number and serial number on the Flex Cardio device, which is on the bottom right corner of the back of the device. Actions to be taken by the customer/user until Philips contacts the customer to arrange for a software update to address this issue, is if the IBP numeric values become frozen and do not update in real-time, opening the Waveform Setup Dialog and then immediately pressing the OK button will restore the proper display of the IBP numeric values. Alternative methods of performing the functions listed in the Problem Description section that do not cause the problem to occur include the following: 1. Change an IBP Site Label using the Pressure Setup Dialog or a macro; 2. Activate or deactivate the mean pressure waveform display option (On/Off) by using the Waveform Setup Dialog, the Pressure Setup Dialog or a macro; 3. Activate or deactivate the waveform hide option (On/Off) using the Waveform Setup Dialog, the Pressure Setup Dialog, the keyboard or a macro; 4. Activate or deactivate the pressure waveform filter option (On/Off) using a macro. A Philips representative will contact each customer regarding their affected Flex Cardio devices. A software update will be required on both the affected Flex Cardio device and on the host workstations driver. These actions will be implemented free of charge by Philips.

FDA

O-SPOT REF 791, 791S Adhere to skin for marking and identifying moles in mammography.

There is a potential for imaging artifact to mimic calcifications on patient's mammograms

On June 14, 2018, Beekley Medical issued an "URGENT MEDICAL DEVICE RECALL" notice dated June 13, 2018 to customers. The notice described the products, problem and actions to be taken. Customers are advised to do the following: - Inspect stock and quarantine affected product - complete and return the acknowledgement response form via fax to: 1-800-735-1234 or mail. - contact firm if inventory of affected product remains with customers at 1-800-233-5539 between 8:00am - 5:30pm EDT. If you have any issues occurred or for further question, contact Quality Assurance & Regulatory Manager at 1-800-233-5539x434.

FDA

Heron toilet and bath chair Model 880505-01 & 880505-11

Replacement campaign on all units that have a gas spring head in aluminum.

The firm sent a letter ask you to replace the gas spring heads on these Herons the replacement kits are being sent to your attention. Please carry out the required retrofitting yourself or forward the kits and instructions to those that maintain the product on your behalf. This replacement must take place as soon as you possibly can and no later than 31-12-2017 and the reporting form documenting this must be filled in for each unit and returned to R82 A/S.

FDA

3.2mm Proximal Reamer/Cannulated Drill provided as part of the Fibulock Fibular Nail System Pack. FibuLock Nail Procedure Pack: Contains instrumentation for the implantation of the FibuLock implant. Contents include: Actuation Driver, 6.2mm Reamer, 3.2mm Reamer, 2 mm Drill, Spade Tip Guide Wire, and 1.6mm x 12 K-Wire.

There is potential to break during use.

Arthrex sent letters to their consignees with the following instructions. 1. Immediately discontinue use, sale, and distribution of the affected product. 2. Please contact Arthrex Product Surveillance at 866-267-9138 or complaints@arthrex.com as soon as possible to arrange for return of the affected product. Our Product Surveillance Specialists can provide assistance regarding alternative solutions and are available to answer questions regarding credit for affected devices in your possession. 3. If you are an Arthrex agency, you do not need to notify any customers as Arthrex will be notifying customers directly. 4. If you are an Arthrex customer, you do not need to notify any patients of this product recall. 5. If you have any questions about this product recall, please contact Arthrex Product Surveillance at 866-267-9138 or complaints@arthrex.com.

FDA

NICO Shepherd' s Hook - Greenberg, sterile. An accessory to NICO s BrainPath device Part Number: NN-8024

Mislabeled hook pouch included a transport/storage label rather than the product label containing key product information such as part number, lot number, and expiration date, etc

On June 19, 2018 , NICO issued letter by mail and/or email letters to consignees that have received hooks from the affected lot. The letter provides instructions to inspect all of its current stock of NN-8024, Greenberg Hooks for the proper pouch label. if an incorrect label is discovered, the hook should be removed from stock and returned to NICO. Questions or need additional information, please contact jay.dittman@.niconeuro.com or 317-660-7118 ext. 104.