Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
Hazard
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
What You Should Do
Remedy
The firm initiated the recall by letter on 05/24/2018. The letter explained the issue and requested the UPS(s) be disconnected from the system until the system could be repaired.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders s
- Brand
- GE Healthcare, LLC
- UPC
- GTIN 00840682124393 Serial Numbers: DDRW34032, DDRX34038, DDRX34041, DDRX34045
Units Affected
4 units