Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
Hazard
A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
What You Should Do
Remedy
The firm initiated their recall by email on 07/28/2017. The firm directed the consignee to repair the units on site and destroy the affected parts.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Carina Mobile Lifts, Models:
a) Carina350EM, REF 6060011
b) Carina350EML, REF 60600012
c) Carina350EE, REF 60600009
- Brand
- Handicare AB Maskinvagen
- UPC
- a) Carina350EM, REF 6060011, Serial numbers: 661102000 - 661102160 b) Carina350EML, REF 60600012, Serial numbers: 661200500 - 661200584 c) Carina350EE, REF 60600009, Serial numbers: 660900500 -660900571
Units Affected
9 units