FDA Device

TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00

Hazard

A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.

What You Should Do

Remedy

Certified letters were sent to all affected customers on 07/14/2017, to offer a replacement of the battery charger.

Products

TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
Brand
Mindray DS USA, Inc. dba Mindray North America
UPC
All battery chargers with part number 115-030108-00

Units Affected

203