FDA Device

Dash 3000 Patient Monitoring System is used to monitor physiologic parameter data on adult, pediatric and neonatal patients.

Hazard

The patient monitors may simultaneously restart as designed if all are connected to the same network and a network overload occurs for a prolonged time.

What You Should Do

Remedy

The recalling firm issued letters dated 8/31/2018 via FedEx on 8/31/2018 informing the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Dash and/or Solar patient monitor on the network does not resume normal functionality in approximately 90 seconds.

Products

Dash 3000 Patient Monitoring System is used to monitor physiologic parameter data on adult, pediatric and neonatal patients.
Brand
GE Healthcare, LLC
UPC
All monitors

Units Affected

Approximately 182,640 monitoring systems for Dash 3000/4000/5000