Monaco Radiation Treatment Planning System (RTP) System
Hazard
Monaco is using the incorrect energy when optimizing and calculating dose.
What You Should Do
Remedy
An Important Field Safety Notification letter is being emailed to affected customers beginning 11/22/2019. The letter identifies affected product, provides a description of the problem, and recommends user not use multiple energies within the same plan until a software fix has been applied to the device. A response form was asked to be returned. Affected customers are advised to contact local Elekta service teams for questions.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Monaco Radiation Treatment Planning System (RTP) System
- Brand
- Elekta Inc
- UPC
- Software Builds: 5.40.00, 5.50.00 and 5.51.00. UDI GTIN: (01)00858164002190(10) 5.40.00, (01)00858164002244(10) 5.50.00, and (01)00858164002268(10) 5.51.00
Units Affected
76 (7 US, 69 OUS)