TosohAutomated Enzyme Immunoassay Analyzer AIA-900, Product Catalog No. 022930. Product Usage: The AIA-900 Automated Enzyme Immunoassay System is an automated analyzer for in vitro diagnostic use that measures a variety of substances in serum, plasma, and or urine using the AIA-PACK reagent series.
Hazard
This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.
What You Should Do
Remedy
Customers were sent a recall notification by email on July 11, 2017 informing them that their local Field Service Engineer would contact them to make arrangements to proactively replace the affected parts. Follow up notifications will be by Federal Express with verification of delivery.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
TosohAutomated Enzyme Immunoassay Analyzer AIA-900, Product Catalog No. 022930.
Product Usage:
The AIA-900 Automated Enzyme Immunoassay System is an automated analyzer for in vitro diagnostic use that
- Brand
- Tosoh Bioscience Inc
- UPC
- All Photo Interrupter sensors that are current in AIA-900 analyzers with serial number(S/N) 10000103 to S/N 11147003.
Units Affected
565 total devices U.S.