Axis-Shield Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT, Part Number FHBC100 - Product Usage: The device can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria. The device is intended for Professional use only.
Hazard
There is a potential performance issue as detected by an under-recovery of sample results during routine testing.
What You Should Do
Urgent Medical Device Correction notification letters dated 9/16/19 were sent to customers. Necessary Actions Immediately discontinue use of the affected lots listed and switch to the alternate material lot(s) you may have available in your laboratory. Immediately destroy any remaining inventory of the affected lots listed according to your laboratory procedures. Complete and return the accompanying verification form Immediately order replacement material(s) if you do not have unaffected material available. Please retain this letter for your laboratory records.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Axis-Shield Diagnostics, Ltd. 17 Luna Place Dundee United Kingdom
- UPC
- UDI: 05055845400282; Lot Numbers 902924530, 902925703
Units Affected
740 total