Carousel MAS1025-01, part of the Brain Surgical Accessories Intra-operative Kit (KIT0295), accessory used with the Renaissance System (TPL0038). Indicated for precise positioning of surgical instruments or spinal implants during general spinal and brain surgery. It is intended to be used whenever the clinician and/or patient benefits from generated 3D imaging of, high contrast objects.

Hazard

In some situations the carousel position pin (locker) may cause the carousel upper plate to pop out and the carousel may then inadvertently move.

What You Should Do

Remedy

All Mazor Brain Clinical Sales Representatives (support during clinical cases) were notified of the issue on 11/26/2014 via email as well as received training.

Products

Carousel MAS1025-01, part of the Brain Surgical Accessories Intra-operative Kit (KIT0295), accessory used with the Renaissance System (TPL0038). Indicated for precise positioning of surgical instrume
Brand
Mazor Robotics Ltd
UPC
Affected products: all Carousels under Level 01 were affected and therefore are applicable to this issue.   The related lots include: WP065611, WP066389, WP066411, WP066412, WP066420.   Explanation of the coding system: a part number (e.g. MAS1025-01) is composed of a generic part number related to the part (e.g. MAS1025) and the level of the part (e.g. 01).

Units Affected

5 units