FDA Device

Calix P PEEK Lumbar System, PLIF Rasp

Hazard

The trials and rasps used to prepare the surgical site for placement of the implant may become detached from the inserter assembly part and become lodged in the intervertebral space.

What You Should Do

Remedy

Consignees will be notified via email and courier and will be instructed to return the affected product. Upon receipt of the recalled trays, the trials and rasps will be removed from the tray and quarantined for evenutal destruction.

Products

Calix P PEEK Lumbar System, PLIF Rasp
Brand
X Spine Systems Inc
UPC
Size, 22mm x 10mm x 6mm, Part #:, X034-0990, Lot #:, 015641-2-1, 015641-2-1R, 015857-24-1, 020316, 020329, 020746, 19894, UP1009, UPI002, UPI005, UPI008 & Size, 22mm x 10mm x 10mm, Part #:, X034-0994, Lot #:, 015641-3-1, 015641-3-1R,015857-25-1, 020317, 020330, 020747, UPI001, UPI004, UPI007.

Units Affected

182 units