CARDIOSAVE Hybrid IABP and CARDIOSAVE Rescue IABP affected part number is 0998-:XX-0800-XX Li-Ion Battery affected part number is 0146-00-0097 Product Usage: The CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons
Hazard
Maquet has received 3 confirmed complaints related to the Li-lon Battery used with the Maquet CARDIOSAVE Hybrid IABP and Maquet CARDIOSAVE Rescue IABP of the battery venting after being accidentally dropped.
What You Should Do
Remedy
Maquet contacted their customers via letter on 12/23/16 and additional customers identified were notified on 1/10/17 and issued a recall letter/return response form. The letter identified the affected product problem and actions to be taken. Customers were instructed to complete and return the response form via fax to 1-973-629-1518 or email to cardiosavebattery@getinge.com
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
CARDIOSAVE Hybrid IABP and CARDIOSAVE Rescue IABP affected part number is 0998-:XX-0800-XX
Li-Ion Battery affected part number is 0146-00-0097
Product Usage:
The CARDIOSAVE Hybrid Intra-Aortic Ball
- Brand
- Maquet Datascope Corp Cardiac Assist Division
- UPC
- part number 0146-00-0097
Units Affected
8000