FDA Device

Encompass Hip Stem, Plasma Spray, 12/14 Taper, Standard, REF 108-XXXX, Sterile R, Ortho Development, Draper, Utah 84020. Intended for use in total hip arthroscopy. Device fixation is achieved by an optimal, press-fit in the medullary canal, which maximizes contact between the stem and the bone.

Hazard

Tests on PVC packaging material resulted in alert levels for cytotoxicity.

What You Should Do

Remedy

Ortho Development Corporation notified consignees by letter on May 11, 2011. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to return all remaining affected units to Ortho. Replacement product was to be shipped by May 31.

Products

Encompass Hip Stem, Plasma Spray, 12/14 Taper, Standard, REF 108-XXXX, Sterile R, Ortho Development, Draper, Utah 84020. Intended for use in total hip arthroscopy. Device fixation is achieved by an o
Brand
Ortho Development Corporation
UPC
REF Number 108- 1017, 1019, 1217, 1219, 1221, 1223, 1419, 1421, 1423, 1425, 1623, 1625, 1627, 1825, 1827. All Lot numbers.

Units Affected

113 units