Medtronic HMS PLUS, Hemostasis Management System. Medtronic, Inc., Cardiovascular Division, Moundsview, MN 55112. A microprocessor based, multichannel clot timing instrument with automated syringe handling for pipetting blood into single use cartridges. It performs in vitro heparin sensitivity evaluations, heparin assays, activated clotting times and platelet function evaluations.

Hazard

The heparin assay controls may run longer than normal and in some cases, fail to give an acceptable result.

What You Should Do

Remedy

Medtronic issued an "Urgent Medical Device Notification" letter dated March 8, 2010 addressed to "Dear Valued Customer" and was sent to three locations at each customer site: Risk Management, Cardiovascular Surgery and Clinical Laboratory-Point of Care. The letter described the product and the issue, provided instructions for control and patient testing. The customer was requested to return an enclosed certificate as proof of notification. For further information, contact your Medtronic Sales Representative or call 1-763-391-9915.

Products

Medtronic HMS PLUS, Hemostasis Management System. Medtronic, Inc., Cardiovascular Division, Moundsview, MN 55112. A microprocessor based, multichannel clot timing instrument with automated syringe h
Brand
Medtronic Inc. Cardiac Rhythm Disease Managment
UPC
HMS Plus Catalog Number: 30514;  Heparin Assay Controls: 306-01POR, 306-02POR, 306-03POR, 306-04POR, 306-05POR and 306-09POR.

Units Affected

508 US, 36 OUS