Medtronic DBS" 3387S-40 Lead Kit for Deep Brain Stimulation , length: 40 cm. Contents of the inner packages are Sterile and Non-Pyrogenic. Method of Sterilization: Ethylene Ox ide, Rx Only, Activa Therapy, Model 3387 Lead Kit Manufactured at : Medtronic, Inc., Villalba, Puerto Rico, STIMLOC Burr Hole Kit: Manufacturer: Medtronic Image-Guided Neurologics, Melbourne, FL 32935-3145 Implant manual: DBS Leads: Proximal 40 mm, Distal 12.0 mm, Medtronic, Inc. Medtronic Active Therapy includes Activ

Hazard

Medtronic has received reports of DBS" leads being damaged at the proximal connector end of the lead when the lead cap is used in the implant procedure. The lead damage is the result of excessive torsion and / or tensile lading of the proximal connector end of the lead. Data also suggest that excessive torque while tightening or loosening the setscrews may twist the setscrew connector block and d

What You Should Do

Remedy

Consignees were sent a "Medtronic Urgent: Medical Device Correction" letter 8/2008. The letter addressed to Healthcare Professionals, described the product involved, explanation of the issue, recommendations for Implant and included an attachment "Updated Instructions for Using the Lead Cap". Contact Medtronic Neuromodulation at 1-800-707-0933 for assistance.

Products

Medtronic DBS" 3387S-40 Lead Kit for Deep Brain Stimulation , length: 40 cm. Contents of the inner packages are Sterile and Non-Pyrogenic. Method of Sterilization: Ethylene Ox ide, Rx Only, Activa Th
Brand
Medtronic Neuromodulation
UPC
PIN # 3387S0000V

Units Affected

8289