EasyRA Cuvette Segment 1, P/N 007000-001 (In Vitro Diagnostic ) EasyRA Cuvette Segments included in the following Kits: EasyRA Cuvettes, REF 10700-4 Startup Kit - Chemistry, REF 10990 Startup Kit - Chemistry & ISE, REF 10991
Hazard
Chemistry analyzers may report Calcium results below the sample's actual concentration. Reported results may be ten to fifteen percent lower than expected sample results. The defect is a shortened path length at specific cuvette positions of those segments manufactured using molding cavity 1. The shortened path length was determined by measuring the cuvette dimensions with a Co-ordinate Measuri
What You Should Do
Medica issued an Urgent Product Recall letter dated March 29, 2010 to customers. Users were requested to cease using and discard the affected product, and complete and return a Response Form in order to obtain replacement product. Customers can contact Medica Technical Support at 877-777-5895.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Medica Corporation
- UPC
- Lot Numbers: 2009/A2810, 2009/A2910, 2009/A3010, 2010/A2001, 2010/A21 01, 2010/A1901, 2009/A0211, 2009/A0311
Units Affected
14,926