ACIST Automated Manifold (Transducer) Kit, Model BT2000, SKU #014613 (used with ACIST Contrast Management Systems (CMS2000) and E2000 Voyager Units)
Hazard
Some of the automated manifold's 2.5 inch high-pressure lines for radiopaque contrast media, used in angiographic procedures, have separated where they are bonded to the manifold body of the device allowing contrast media to leak.
What You Should Do
Remedy
The consignees were sent letters dated 10/2/03 which requested that the product be returned.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
ACIST Automated Manifold (Transducer) Kit, Model BT2000, SKU #014613 (used with ACIST Contrast Management Systems (CMS2000) and E2000 Voyager Units)
- Brand
- Acist Medical Systems
- UPC
- Lot Numbers 0933A and 1003I
Units Affected
7120 kits