KODAK DirectView DR 7100 System, Catalog # 133-6049. This is a digital radiology system. The major components are: operator console; digital imaging capture device; tilting table with floating top; ceiling-mounted x-ray tube; x-ray generator timing and distribution unit.
Hazard
Potential weld defect in the Overhead Tube Support assembly which connects the tube assembly to the telescoping arm. This could cause the tube assembly to fall, injuring anyone near the unit.
What You Should Do
Remedy
Letters dated 10/20/2005 were hand carried by Kodak Field Service Reps. Kodak Field Engineers inspected units and found the two defective units (both in China).
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
KODAK DirectView DR 7100 System, Catalog # 133-6049. This is a digital radiology system. The major components are: operator console; digital imaging capture device; tilting table with floating top;
- Brand
- Eastman Kodak Co
- UPC
- s/n DR71000121 and DR71000123.
Units Affected