Zimmer brand Sirus Intramedullary femur nail system step reamer 0 mm, 6.5/4.5 mm cannulated, for cervical screws, item number 02.00020.039.
Hazard
The calibrations on the reamers may lead to inaccurate reaming depth when used in accordance with the published surgical technique.
What You Should Do
Remedy
Consignees were notified via recall letter dated 9/30/05.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Zimmer brand Sirus Intramedullary femur nail system step reamer 0 mm, 6.5/4.5 mm cannulated, for cervical screws, item number 02.00020.039.
- Brand
- Zimmer Inc.
- UPC
- Lots 480103, 480104 and 487665.
Units Affected
27