Datex-Ohmeda S/5 and AS/3 Anesthesia Delivery Units (The S/5 units are a later version of the AS/3 units. The conversion from AS3 to S5 occurred during the summer of 2000. This change was mainly a product name change, with a slight color variation to the shade of white being used externally on the machine. There were no design or feature and function modifications when the name change occurred.)
Hazard
Anesthetic agents could be delivered at concentrations above the concentration set on the S/5 and AS/3 Anesthesia Delivery Units (ADU) having serial numbers within a specified range.
What You Should Do
Remedy
Recall letters dated February 03, 2004 stated that a firm representative would contact the consignee to schedule a visit to replace software and hardware to correct the problem.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Datex-Ohmeda S/5 and AS/3 Anesthesia Delivery Units
(The S/5 units are a later version of the AS/3 units. The conversion from AS3 to S5 occurred during the summer of 2000. This change was mainly
- Brand
- Datex-Ohmeda Inc One Ohmeda
- UPC
- The affected units in the United States have serial numbers between 40013635 and 40162379.
Units Affected
1617 units in the United States