Roche brand ISE SnapPak for use only in the AVL 9120/9130 Electrolyte Analyzer Assembled and distributed by Roche Diagnostics Corp., Indpls., IN; Ref/No. 03112276180; Part number BP5016.
Hazard
A defective seam in the two chamber pouch may result in the mixing of reagents, which would either result in an inability to calibrate the analyzer or would bias patient results.
What You Should Do
Remedy
Own name distributor consignees were notified on or about 1/3/05 via letter dated 12/21/04, and were requested to discontinue shipment of the product, notify all of their customers of the recall and to discard product. Roche consignees were notified via letter dated 12/21/04, requested to notify their customers if they had further distributed the product, and instructed to discard all inventory of product.