ACCU-CHEK Inform meter; catalog No. 3035123.
Hazard
A firmware bug may allow test results from other systems, recorded into the INFORM system via the OTE functionality, to be reversed from positive to negative, and vice versa.
What You Should Do
Remedy
Consignees were notified via a letter dated 11/19/04, which instructed them to discontinue using the OTE functionality.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
ACCU-CHEK Inform meter; catalog No. 3035123.
- Brand
- Roche Diagnostics Corp.
- UPC
- All units with firmware version 2.3 that have the OTE (other test entry) functionality enabled.
Units Affected
4697