Roche COBAS INTEGRA HBA1c Kits for use with COBAS INTEGRA 400, 400+, 700 and 800 analyzers; catalog 20753521322.
Hazard
The product is not uniform across the cassettes of some lots, which will result in an inconsistent bias of patient results.
What You Should Do
Remedy
Recall letters dated 2/15/05 were sent to each user instructing customers to run controls before the use of each cassette or kit in a lot, instead of once per lot as previously instructed, and to notify the physician at the facility.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Roche COBAS INTEGRA HBA1c Kits for use with COBAS INTEGRA 400, 400+, 700 and 800 analyzers; catalog 20753521322.
- Brand
- Roche Diagnostics Corp.
- UPC
- Lots 660587, 659867, 657295, 657344, 658498, 656157, 655052, 653288, 653424, 652153, 650988, 650430, 650119, 649291, 647186, 646336, 644962, 643398 and 642104.
Units Affected
347,000 cassettes