TDx/TDxFLx C-Reactive Protein (CRP) Reagents, list 9550-60; Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064
Hazard
The reagent package inserts for the products contain incorrect information in the Specific Performance Characteristics section. Accuracy by Correlation table
What You Should Do
Field Correction recall by Important Product Information letter dated 12/5/03, sent via Federal Express Priority on the same date to all customers receiving the affected reagents. The accounts were informed of the incorrect values for the Specific Performance Characteristics listed in the package inserts of the reagent kits. The letters contained a table listing the incorrect values and the correct values for each of the products, and the users were instructed to replace their product inserts with the enclosed revised package inserts for all product specific reagent kits in their laboratory.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Abbott Laboratories HPD/ADD/GPRD
- UPC
- All lots with insert version 66-9771/R3. The corrected insert is version 34-2897/R4.
Units Affected
1,644 kits