BD Kiestra InoqulA Software Version 20.3.IVD automated specimen processing Catalog Number: 447202
Hazard
An anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate
What You Should Do
BD initiated a field correction via letter on July 26, 2019 and BD Technical Service has completed the correction (software upgrade) for all six (6) affected US customers. Letter identifies issue, health risk and customer action to take. BD is in the process of evaluating the log files of customers using the InoqulATM / InoqulA+TM software version 20.3 to determine if any users potentially experienced mismatch between a specimen and plate. BD will contact all customers using InoqulATM / InoqulA+TM software version 20.3 to provide the outcome of this evaluation. Contact: BD Customer/Technical Support 800-638-8663 Monday Friday between 7:00am and 7:00pm (EST) in the United States. For customers outside the US, contact your local BD representative or distributor.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Becton Dickinson & Co.
- UPC
- Catalog Number: 447202 Catalog Number/UDI INO-000287 00382904472023 INO-000286 00382904472023 INO-000269 00382904472023 INO-000281 00382904472023
Units Affected
4 units