December 2017
SynchroMed EL, models 8626-10, 8626L-10, 8626-18, 8626L-18, 8627-10, 8627L-10, 8627-18, 8627L-18. implantable infusion pump. The pump is part of an implanted infusion system that consists of a SynchroMed EL pump and a catheter. The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion of drugs or fluids. The SynchroMed EL Infusion System with a 10 mL reservoir volume is indicated for use in the smaller patient who has insufficient body mass to s
Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.
The firm Medtronic, sent an "Urgent medical Device Safety Notification" dated November 2012, to its customers beginning November 09, 2012. The letter described the product, problem and actions to be taken. The customers were instructed to follow the recommendations provided in the letter and complete and return the PHYSICIAN REPLY FORM via mail at Medtronic Neuromodulation Quality Assurance, 7000 Central Ave NE, Mailstop RCC150, Minneapolis, MN 55432; fax to: 800-897-3899 or email to: neuro.quality@medtronic.com. If you have questions, please contact Medtronic Neuromodulation Technical Services at 1-800-707-0933 weekdays 7am - 6pm CST. UPDATE 2/1/2016: Medtronic began notifying consignees 2/1/2016 via, mail, telephone and/or in person visit by a Medtronic Field Representative. The Medtronic Field Representatives will retrieve devices in the US and provide an Urgent Medical Device Removal letter. The letter states that device design has been updated and the Medtronic Field Representative has retrieved any devices with a Use By date on or before 5/14/2017. An account specific Customer Confirmation Form was to be completed by either the Medtronic Field Representative or the consignee. Consignees with questions can contatct their Medtronic Field Representative, or Customer Service at 1-888-854-0978.
Microalbumin Reagent, Catalog No. 475100, Product Labeling A18520AL MA reagent, when used in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems MA Calibrator, is intended for quantitative determination of Albumin concentration in human urine
Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.
Beckman Coulter sent an Urgent Medical Device Recall letter dated April 24, 2017, to affected customers to inform them that Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1. Internal testing has confirmed that MA is recovering within the published insert range for Beckman Coulter Urine Protein Control Level 2. Customers are informed of the impact, actions to be taken, and resolution. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to Customer Support Center via http://www.beckmancoulter.com/customersupport/support, call 1-800-854-3633 in the United States and Canada, and outside US and Canada they are instructed to contact their local Beckman Coulter Representative.
Centricity Laboratory Core Lab System 4.1 The Centricity Laboratory System is intended to be an information system designed to support the clinical and administrative activities associated with the provision and utilization of clinical laboratory services and facilities, e.g., the storing and delivering of analytical results. It is a specially designed data program application (software), which is supplied for installation in existing mainframe or decentralized computers or a computer network.
The Centricity Laboratory removes free-text notes on a master panel when the user deletes the slave panel. The deletion of a slave panel will (auto slash) the corresponding test (item) results on the master panel. Additionally, the displayed and HL7 results show neither the appropriate abnormal flag value nor the reference range values, when using large_num.g >7 to report numeric results.
GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated Nov 20, 2012 to all affected customers. The letter described the product, problems and actions to be taken. The customers were instructed to follow the safety instructions and to ensure that all potential users in their facility are made aware of this safety notification and the recommended actions. The software has been modified to correct these defects. The modified software will be released and installed in user sites through GE Healthcare FMI #86071. Customers were instructed to contact their GE Service Representative with any questions regarding this issue. For other questions, contact GE Helpdesk: North America; (905) 479-1249 or 1-888-778-3375; email: tgg.nahelpdesk@med.ge.com Europe; +44 1603 877595; email: tgg.ukhelpdesk@med.ge.com Asia Pacific; +61 3 5327 4000; email: tgg.ozhelpdesk@med.ge.com. For questions regarding this recall call 847-277-5000.
Baxter Interlink System Buretrol Solution Set with 150 mL Burette (Ball-Valve Drip Chamber); a sterile Rx IV fluid pathway device. Product Usage: The device is used for the administration of fluids from a container into the patient's vascular system through a vascular access device.
Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
Baxter Healthcare Corp. sent an Urgent Product Recall letter dated September 7, 2012, via first class mail to all affected customers, informing them of the problem with the ball-valve, and instructing them to discontinue use of the affected product and to contact Baxter for instructions to return the affected product.. The customers were also requested to complete the attached customer reply form, confirming their receipt of the letter, and fax it to Baxter at the number provided on the form. Any questions regarding the communication were directed to Medical Information Services at Baxter at 1-800-933-0303.
Vantage Panoramic X-Ray System used to perform dental and orthodontic radiological imaging The Vantage Panoramic X-Ray System is a class II extra oral X-Ray system used to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometric radiographic examinations for use in orthodontic treatment planning and evalu
Failure of the primary collimator may result in unintentional movement of the device. The unintentional movement will not result in the generation of an error message.
MidMark sent an Urgent Medical Device Field Correction letter dated September 4, 2014, to all affected customers. The letter included instructions for distributors to: 1) encourage end users to contact Midmark to arrange the field correction after the final solution becomes available; 2) if a partial image is reported to have occurred with a machine, the machine should be removed from use and the end user should initiate a service request per the normal procedure; 3) quarantine any affected machines in the distributor's inventory and contact Midmark Technical Support within 10 business days to provide the Serial numbers to on-site servicing of the affected machines can be scheduled. Distributors with questions about this market action can contact Midmark Technical Support by calling 888-294-3800 x 3 or by e-mail at imagingtechnsupport@midmark.com.
8" x 4" 16-ply X-Ray Gauze, hospital fold, banded in 10's; 200 pieces per bag, 10 bags per case; Distributed By: International Medsurg Connection, Schaumburg, IL USA 1-847-619-9926; Made in China; item IMSGZX8416-10 For use to control bleeding and absorb fluid.
The lots of X-Ray Gauze are contaminated with foreign material, including remnants of cotton seeds and lint.
International Medsurg Connection notified their direct accounts via telephone, e-mail and/or fax on April 9, 2012, informing them of the foreign material contamination of the gauze, and requesting them to quarantine all affected lots in inventory, cease distribution of the affected product, and notify any customers at the retail level to remove and discard these products. Any questions were directed to (847) 619-9926.
Medtronic PAK NEEDLE BEVELED AND TROCAR TIPS, REF 8670009, QTY 1 EA, STERILE R, Rx only
Product packaging may incorrectly include two bevel-tipped needles or two trocar-tipped needles instead of one diamond-tipped needle and one bevel-tipped needle.
Medtronic sent an Urgent : Field Safety Corrective Action dated April 13, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Instructions to Customers: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via email to rs.sdmriskmgt@medtronic.com, or fax to 763-367-8134. 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return and replacement of affected product. We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please contact our Post-Market Quality team at 901-399-3197. For further questions, please call (901) 344-1435.
Jackson-Pratt Hemaduct Cardiothoracic Silicone Round Drain,19FR. Sterile, Rx Only. For Single Use Only. Jackson-Pratt Wound Drainage Systems are sterile single use devices. The wound drainage system consist of a wound drain and a fluid collection reservoir. The wound drain is a radio-opaque section of tubing which has perforations or ducts. The wound drain is surgically placed in a surgical wound site using a trocar. Blood and fluids are collected through a wound drain into a fluid collectio
Product's seal possibly compromised, potentially compromising the sterility of the package contents. Use of impacted products could result in an increased risk of infection.
CardinalHealth sent an Urgent Product Recall letter dated April 21, 2017, to all affected customers (including response firm). Customers were instructed to inspect recalled lots and return any affected product found. Distributors were instructed to notify their customers. Affected product will be destroyed. To arrange for return of any affected product by contacting Customer Service at the following numbers: Hospital800.964.5227 Distributor800.635.6021 Federal Government800.444.1166 All Other Customers888.444.5440 Customers with questions were instructed to contact Cardinal Health Customer Advocacy at 800-292-9332. For questions regarding this recall call 847-887-4138.
DiaScreen Liquid Urine Controls: Positive and Negative Assayed Liquid Quality Controls for use with Dipsticks in Routine Urinalysis; Product Number D76004 (4x12 mL bottles) consists of: 2x12mL Positive controls [and] 2x12mL Negative controls; Manufactured for ARKRAY USA, Inc.; Minneapolis, MN 55439 Diascreen liquid urine control is a synthetic control designed to react with commercial urine dipsticks to register listed responses on color reagent pads. It is intended for use in a clinical labora
The products may have been stored at temperatures outside their required storage conditions.
Independence Medical (im) sent an Urgent Product Recall letter dated July 16, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to identify and immediately cease the distribution of the recalled products, quarantine the recalled products, and contact the recalling firm at 800-860-8027 for instructions on how to return the recalled products. Consignees who further distributed the recalled products were requested to notify their customers of the recall. Consignees were also instructed to complete the Product Recall Acknowledgement Form and return the document by email to recall@indemed.com indicating receipt of the notification.Customers with questions were instructed to contact the product manufacturer for ProTime Disposable Cuvettes at 800-579-2255, DiaScreen Liquid Urine Controls at 800-818-8877, and Hemoccult II SENSA elite Dispensapak at 800-854-3633. For questions regarding this recall call 800-860-8027.
Wound Pro Apex Negative Pressure Wound Therapy Powered Suction Pump; Accuro Medical Products LLC, 2854 Hitchcock Ave, Downers Grove, IL 60515; part number 84635400182, sold inside the Apex Negative Wound Therapy system, part number 84635400181 Promotes wound healing through drainage and removal of infectious material and other fluids from the wound site using continuous and/or intermittent negative pressure. Types of wounds indicated include diabetic/neuropathic ulcers, pressure ulcers, chronic
Accuro has received reports of some Apex Pumps either failing to operate or malfunctioning, at times without notifying the user of such failure or malfunction on the unit's display screen.
The firm, Accuro, sent an "Urgent: Medical Device Recall" letter dated August 30, 2012 to the customers via e-mail on August 31, 2012. The letter describes the product, problem and actions to be taken. The customers were instructed to remove all devices from use immediately; to immediately complete and return the enclosed Accuro response form via email to; (customer.service@accuromedical.com) or fax to: (1-800-866-0545) . Note: the form still needs to be returned even if you no longer have or use the device. Upon receipt, Customer Service will call the account with instructions to arrange for all devices to be returned to Accuro. General questions were directed to Customer Service at 1-800-669-4757.
CORFLO 300 Enteral Pump; a volumetric rotary peristaltic intermittent enteral feeding pump; CORPAK MedSystems, Wheeling, IL 60090; catalog/reorder #20-0500. Intended to deliver formula to patients requiring tube feedings.
Corpak is removing the Corflo 300 Enteral Pump from the market due to the possibility that with fluid ingress or other damage, the keypad may not function, and the possibility of motor mount failure.
Corpak MedSystems sent Recall Notification letters dated September 19, 2011 to all CorFlo 300 Enteral Pump customers on the same date. The customers were informed that there is a possibility that with fluid ingress the keypad may become inoperable, and that there is a possibility of motor mount failure. Either event could possibly affect the rate at which formula is delivered to the patient, and fluid ingress could cause a potential of audible alarm failure. Corpak determined that the removal of the pumps from the market was the best option for their customers, and requested the return of all CorFlo 300 pumps to Corpak by no later than December 1, 2011. The customers were requested to contact Corpak at 1-800-323-6305 to arrange for the return of the pumps, and to complete and return the enclosed acknowledgement form, indicating the number of pumps on hand, via fax at 847-541-9526 or e-mail to questions@corpak.com. Dealers/distributors were requested to notify their customers of the recall.
Reservoir Bag for WalkMed Infusion Pump Part Number 204820 Catalog Number IPR-86/10 65 mL and Reservoir Bag for WalkMed Infusion Pump Part Number 204821 Catalog Number IPR-150/10 150 mL The WalkMed Reservoir Bag is designed for use with the WalkMed 350VL Ambulatory pump and is intended to be used as a reservoir and fluid path for administering intravenous, subcutaneous, arterial, enteral, and epidural infusions of antibiotics, analgesics, chemotherapeutic agents, and other medications or fluids.
There have been reported incidents of fluid leaks at the luer connection. Leaks have been confirmed to originate from, or near, the reservoirs clearcolored female luer connecter.
WalkMed sent an Urgent Medical Device Recall letter dated September 27, 2017, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: discontinue distribution of any affected products. Contact any customers to whom these products were dispensed and inform the customers accordingly based on the following items. Wear gloves and other personal protective equipment as needed to avoid contact with leaking infusion fluids when handling the product. Collect and return affected products using proper biohazardous material handling where applicable. Complete the enclosed Acknowledgement Form to record the return or destruction of the product and return to WalkMed LLC. Customers with questions were instructed to call 720-351-4966.
IntelliVue Multi Measurement Server, IntelliVue Multi Measurement Server SLCP, IntelliVue Multi Measurement Server X2, IntelliVue Patient Monitor Models MP2 and MP5 Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment.
Distorted ST-Segment when using 12-Lead ECG Monitoring with Philips IntelliVue Patient Monitors
Philips sent an Urgent Medical Device Correction letter to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Philips will provide software update and is provding interim instructions for the operator. For questions contact your local Philips representative.
LIAISON Estradiol II Gen, 310680, 310680-CN The DiaSorin LIAISON Estradiol II Gen assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of estradiol in human serum. Assay results should be used in conjunction with other clinical and laboratory data to assist in the diagnosis and management of conditions involving an excess or deficiency of estradiol. The test has to be performed on the LIAISON Analyzer Family.
DiaSorin is issuing an urgent medical device correction for the LIAISON Estradiol II Gen assay. Due to the risk of a recently identified cross reactivity, the LIAISON Estradiol II Gena assay should not be used to test patients being treated with fulvestrant (Faslodex).
DiaSorin Inc. sent an Customer Notification letter dated July 13, 2016, to all affected customers. The letter described the problem and the product involved in the recall. The letter listed the "Recommended Actions" and requested customers to complete and return the response portion. For questions contact Product Support at 1-800-328-1482.
The Vantage Digital Panoramic System delivers high-value features standard on every system: - Five pre-programmed exam settings, including bitewing, minimize exam set-up time and allow routine diagnostic exams to be performed extraorally for improved clinical efficiency and patient experience. - Progeny Clarity Enhanced Filter Set provides superior diagnostic imaging quality - Exclusive Focal Trough Alignment reminder assists in consistent patient positioning - Progeny Imaging software included
An error was identified in the Real Time Controller (RTC) firmware versions 3.0, 3.1, 3.2, and 3.3 used in Vantage Panoramic X-Ray systems. The error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. failure to stop").
An URGENT MEDICAL DEVICE FIELD CORRECTION letter dated September 20, 2013 was sent to all consignees. Letters sent to distributors include instructions on how to identify the defective units, to quarantine the machine(s), and to contact Midmark at 847-415-9739 or via- e-mail at abavuso@midmark.com. Midmark will either schedule an on-site field correction or arrange for the replacement of the affected device(s) depending on the customer's preference. Midmark further requested the distributors provide Midmark with the names, addresses, phone numbers, and e-mail addresses of any customers that purchased any defective units. Field corrections include the upgrading of RTC firmware to revision 3.4 on the defective units and will be performed as no additional charge by Midmark personnel. Letter sent to practitioners include instructions on how to identify the defective units, instructions on how to temporarily reduce the likelihood of the unintentional movement, and to contact Midmark at 847-415-9739 or via- e-mail at abavuso@midmark.com. Midmark will schedule an on-site field correction by upgrading the firmware to revision 3.4.
1) The FEMTO LDV Z2 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery or other treatment requiring lamellar resection of the cornea at constant depth relative to the corneal surface. 2) The FEMTO LDV Z4 Femtosecond Surgical Laser is an opthalmic surgical laser indicated for use in the creation of the corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of ri
A defect in the controller board can lead to the condenser to the video monitor becoming too warm. This could potentially cause the video monitor to shut off.
A recall notification letter dated April 16, 2013 was mailed to direct accounts. The letter notified the direct accounts of the problem and advised them that a technician would be arriving at their facility during the week of June 2, 2013 to replace the defective controller board. The technician will also perform preventative maintenance on the device at the same time. Direct accounts are advised to contact Ziemer USA, Inc. at 618-462-9301 if they have any questions about this issue.
Convertors Robotic Drape with Attached Leggings; Product is packaged as Single Sterile Drape/8 per Case
Product 29474 LAVH drape was packaged as 29574 Robotic Drape with Leggings.
Cardinal Health sent an Urgent: Product Recall letter dated July 28, 2017. Distributors are requested to notify their customers of the action. An acknowledgement form is included as part of the notification. Responses are tracked. All returned product will be quarantined. All inventory recovered from the field will be destroyed. For further questions, please call (800) 292-9332.
The ZYTO Tower is the input device to program the software with various virtual items.
ZYTO Technologies Inc. Announces a Voluntary Recall of the ZYTO Tower and ZYTO Select and Elite Software Due to Claims Exceeding the 510(k) Clearance.
Consignees were notified on 11/23/2015 via letter and email of an update to the ZYTO Select and Elite Software that would disable all tower functionality..
Pentax Video Bronchoscope Designed to be used with a Pentax Video Processor (including light source), documentation equipment, video monitor, endo-therapy accessories (such as biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the airways and tracheobronchial tree.
Pentax America Inc. is recalling various bronchoscopes which contain an incorrect Operation and Reprocessing IFUS.
Pentax Medical sent a U.S. Urgent Field Correction notice to all affected customers with a recall letter/return response form on May 31, 2016 via Fed Ex. Customers with questions were instructed to call 800-431-5880, ext. 2064.
Bronchoscope Intended to provide optical visualization of and therapeutic access to the Airway and Bronchial Tree. This anatomy includes, but is not restricted to, the organs, tissues and subsystems: Nasal Passage, Trachea and Bronchial Tree. The instrument is introduced via the mouth or the nose, as decided by the physician, when indications consistent with the need for procedure are observed in adult and pediatric patient populations.
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Pentax sent an Urgent Field Correction letter dated April 6, 2016 to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were reminded to carefully check that all accessories are intact, that no parts have fallen off and become lodged within the endoscope's instrument/suction channel and to ensure that any therapeutic devices passed through the instrument channel are accounted for after use. If the instrument/suction channel does become blocked or clogged due to the accumulation of debris, an accessory that cannot be removed, or other cause, do NOT attempt to correct the blockage or continue to use the endoscope. Customers were instructed to contact their local PENTAX Medical Service facility to have the endoscope repaired. Customers with questions were instructed to call 800-431-5880, ext 2064.
Mallinckrodt Pharmaceuticals INOmax DSIR Plus delivery system. Model 10007. Delivers INOMAX (nitric oxide for inhalation). The INOmax DSIR uses a dual-channel design to ensure the safe delivery of INOMAX. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and 02 cells) and the user interface including the display
Mallinckrodt discovered that some devices may have been manufactured with an out-of-tolerance manometer which can result in in an inaccurate proportional valve calibration table in an affected device.
Recall letter was sent to customers in a e-mail beginning November 21, 2016 with a had copy sent via FedEx overnight with signature receipt beginning December 5, 2016. Devices will be picked up and replaced by Mallinckrodt. Questions can be directed to 1-877-KNOW-INO (1-877-566-9466).
MesaLabs Conductivity Standard Solution, 14 mS/cm, 32 ounces (950 ml). Product No. : 02.0027, 020510 (carton of 6 of 02.0027). Polyethylene bottle Conductivity/TDS Calibrator solutions are a secondary standard solution used for the calibration of conductivity/TDS cells together with conductivity/TDS measurement instruments. The conductivity/TDS cells and instruments may be indicated for calibrating the conductivity measurement function of hemodialysis machines and water purification equipment
Mesa Laboratories is recalling Conductivity/TDS Calibrator Solution due to failure to meet product specifications.
MesaLabs sent an Device Correction Notice dated May 12, 2017, via certified letter to destroy product by pouring it down the drain and free replacements will be shipped. For further questions, please call (303) 987-8000 ext. 120.
ANGIOGRAPHY PACK; REF: DYNJ26855D; Packaged in Mexico by: Medline Industries, Inc., Mundelein, IL 60060; 1-800-Medline; www.medline.com Surgical kit used to supply tools required to perform angiography medical procedures.
Guidewire .035x150 3MMJ TCFC, item number 88241, has a potential for the coating to flake off of the wire. The guidewire is used in a number of surgical convenience kits assembled and marketed by Medline Industries, Inc.
The firm, Medline Industries, Inc., sent a "RECALL IMMEDIATE ACTION REQUIRED" letter dated August 14, 2013, to its direct accounts/customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately check their stocks for inventory of the affected products, quarantine any affected products, return the product using the enclosed shipping slips for credit-If additional stickers are needed, please call 866-359-1704, and complete and return the enclosed FAX-BACK VERIFICATION FORM listing the quantity of affected product(s) on hand via fax to Medline at 866-767-1290. If you have any questions, please contact 866-359-1704.
Hemotherm CE, model 400CE, Heater-Cooling Device
Improperly rated fuses were used in the manufacture of the device.
Letters were sent to customers and distributors instructing them to immediately cease use and/or further distribution of affected product, remove and discard the affected fuses and replace with the new fuses provided by CSZ, and complete and return the Field Action Response Form.
Inion GRT Tack, drill bit and tack appl., 4x, Sterile. Is indicated for the surgical treatment of periodontal defects to aid in the regeneration and integration of tissue components in guided tissue regeneration procedures, for pre-implant and peri-implant surgery and for covering bone defects and empty sockets.
Aluminum pouch seal was noticed to be defective (slightly open).
Inion sent an Urgent: Medical Device Recall Notification letter dated March 1, 2017. The recall letter asked customers to inspect the aluminum pouch seals, remove any devices with defective seals from sales stock, and to return the product to Inion. The customers were also asked to inform Inion Oy of the outcome of their inspections using a feedback form. For further questions call (954) 659-9224.
ARTISTE" MV System
Software update
Siemens mailed an Customer Information Letter on 02/15/2017 to affected customers to inform them of a software update they plan to implement to reduce risk of collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table, using the SIMTEC-Auto Field Sequence Option.
AMT Bridle - Nasal Tube Retaining System Used with Nasogastric/Nasointestinal (NG/NI) tubes.
Expired expiration dates are listed on the pouches.
Applied Medical Technology Inc. sent an Urgent Medical Device Recall letter dated January 19, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. ACTIONS TO BE TAKEN BY THE CUSTOMER OR USER: Applied Medical Technology requests that customers IMMEDIATELY TAKE THE ACTION and return all devices in inventory subject to this recall. Do not use any device subject to this recall. Complete and return to Applied Medical Technology the enclosed RECALL NOTIFICATION. For further questions please call (440) 717- 4000.
Merge Cardio software. Product Usage: Merge Cardio is an integrated cardiovascular information system classified as a picture archiving and communications system.
Patient information in the header is only found on the first page of the report and not on the subsequent pages.
Merge sent an Urgent Medical Device Recall letter dated January 28, 2016 to affected customers via e-mail or via certified mail if they did not have an e-mail address. The letter identified the affected product, problem and actions to be taken. For questions call Merge Customer Service at (877)741-5369 or support@merge.com.
Smiths Medical CADD Blue-Striped IV Administration Set, REF 21-7036-01, Sterilized using ethylene oxide.
The label of the CADD Blue-Striped Administrative Set is missing the expiration date, which is considered a regulatory compliance issue for Canada.
Customers were sent an Urgent Medical Device Safety Notice, dated 12/9/2016. The letter identified affected product and stated the reason for recall. The letter also provided a list of devices that each consignee ordered, along with the date of manufacture and associated expiration date. Customers were asked to locate devices and compare to the list and determine if affected devices are within expiration date. If the device is expired customers are to dispose of them per their procedures for dealing with expired product. If devices are not expired, they may continue to use them through the expiration date. A response form was asked to be returned within 10 day s of receipt of the letter. Questions can be directed to smithsmedical6185@stericycle.com
CADD Administration Sets with Flow Stop, Product Reorder No: 21-7321-01, 21-7321-24, 21-7322-01, 21-7322-24, 21-7323-24, 21-7324-01, 21-7324-24, 21-7333-24, 21-7336-01, 21-7336-24, 21-7339-01, 21-7339-24, 21-7359-01, 21-7359-24, 21-7383-01, 21-7383-24, 21-7390-01, 21-7390-24, 21-7391-01, 21-7391-24, 21-7394-01, 21-7394-24, 21-7395-24. The CADD Medication Cassette Reservoirs with Flow Stop are a modification to the current CADD Medication Cassette Reservoirs. The reservoirs will incorporate a
Smiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow Stop.
Smiths Medical sent an "Urgent Medical Device Field Correction Action" letter dated April 20, 2016. The letter was addressed to "Clinicians who oversee the use of the CADD Administration Sets with Flow Stop and Distributors thereof". The letter described the Affected devices, Reason for Field Correction Action, Risk to Health, and Instructions to Customers (Distributors and Clinicians). Requested consignees to complete and return the "Field Corrective Action Confirmation Form". For questions regarding this notification, contact Smiths Medical Customer Service Department at 1-866-831-8399.
GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Archiving and Communication System Used as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. Also used as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print and export images when connected w
While merging exams in a test instance, merges carry forward in the production system, when the middle tier is configured to the wrong IP address of the Centricity Exam Manager. When a current and historical exam are opened at the same time, the system asynchronously tries to access the operation specifying how each image should be grouped, causing random image-series grouping errors.
GE sent an Urgent Medical Device Correction letter dated November 15, 2012. The software has been modified to correct these defects. The modified software was released and installed in user sites under GE Healthcare FMI #85196.
The synthetic non-absorbable surgical sutures CARDIOXYL are composed of a polyester [poly(ethylene terephtalate)] braids, with or without PLEDGET of different sizes. CARDIOXYL braids are coated with polysiloxane. CARDIOXYL sutures are green dyed (D&C green # 6, CI 61565) and white (undyed). Some packs are composed of green and white braids to facilitate location during use. CARDIOXYL sutures comply with the requirements of the United States Pharmacopoeia (USP) and the European Pharmacopoeia (EP)
A potential splitting of the surface treatment of the thread ends on certain sutures has recently been identified.
The sole US distributor (MedAlliance Inc.) was notified of the recall via e-mail on July 16, 2014. IMPORTANT AND URGENT - BATCH RECALL PROCEDURE letters dated June 25, 2014 under the Peters Surgical header were sent to all MedAlliance Inc. consignees. The letters included instructions for consignees to contact MedAlliance Inc. by phone (630-933-8661) or by e-mail (agoos@medalliancesolutions.com) for information on what to do with the recalled products. Consignees were also notified that they can contact Peters Surgical directly at qualite@peters-surgical.com for answers to any additional questions. MedAlliance Inc. sent VOLUNTARY PRODUCT RECALL/WITHDRAWAL LETTER[s] dated June 25, 2014 to all consignees. The letters included instructions for consignees to: 1) discontinue use of and segregate recalled product in a secure location for return to MedAlliance Inc., and 2) to complete and return the enclosed RETURNED GOOD AUTHORIZATION (RGA) FORM to MedAlliance Inc. to arrange for the return of the recalled products. Consignees with any questions about the recall can contact Ms. Alaina Goos, Senior Product Manager, by phone at 630-933-8661 x 104 or via e-mail at agoos@medalliancesolutions.com.
MANOS EX (CAT#TY-12-101) Product Usage: Carpal Tunnel Release: as a tool for tissue release of the transverse carpal ligament in patients with Carpal Tunnel Syndrome who fail conservative therapy. Plantar Fasciotomy: as a tool for tissue recession in patients with plantar fasciitis who fail conservative therapy.
The firm failed to notify current customers of IFU update/warning. During a recent FDA inspection, it was discovered that Thayer Intellectual Property, Inc. had revised their IFU to include a warning not to bend, deform or otherwise modify any part of the device. The warning was added as a result of a MAUDE report filed in December 2014 where a piece of the MANOS EX broke during use and was left in the patient's hand. The patient suffered a tendon injury and an additional surgery was required to remove the broken piece of the device and repair the tendon. FDA inspection found that while the firm had updated their IFU, they had not notified customers who had been shipped the device with the older version that did not include the warning. Risk: It is important for MANOS EX users to be aware that bending, deforming, or modifying the device could result in device breakage which could impact patient safety and device performance.
Thayer sent an Urgent Medical Device Correction letters dated November 4, 2016 by Federal Express. The letter identified the affected product, problem and actions to be taken. Customers were asked to check if any of the affected lots were still available, and if so, to change out the IFU with version 5. For questions call 415-923-9375.
Boston Scientific, INGEVITY" MRI Leads, Sterile EO, RxOnly.
Boston Scientific is conducting a recall on Boston Scientific INGEVITY" MRI endocardial pacing leads. There is a possibility that the polyurethane boot at the terminal end of the lead was not securely connected to the lead body.
Boston Scientific sent an "Important Medical Device Information" letter dated November 30, 2016. The letter was addressed to "Dear Doctor". The letter described the problem and the product involved in the recall. The letter also described the "Clinical Considerations, Recommendations, Affected Population & Further Information". For questions contact your local Boston Scientific representative or Technical Services. For further questions, please call ( 651) 582-4000.
EPIQ DIAGNOSTIC ULTRASOUND SYSTEM, Model EPIQ 5C, EPIC 5G, EPIQ 5W, EPIQ 7C, EPIC 7GC, and EPIQ 7W. Diagnostic Ultrasound System for ultrasound imaging in abdominal, cardiac adult, cardiac other (fetal), cardiac pediatric, cerebral vascular, cephalic (adult), cephalic (neonatal), fetal/obstetric, gynecological, intraoperative (vascular), intraoperative (cardiac), musculoskeletal (conventional), musculoskeletal (superficial), other: urology, pediatric, peripheral vessel, small organ (breast, thy
The fasteners securing the control panel assembly to the base of the Philips EPIQ Ultrasound System may loosen over time, which could subsequently lead to the detachment of the entire assembly from the ultrasound system.
The firm, Philips, sent an "Urgent-Medical Device Correction" Philips EPIQ Ultrasound System (MDC 79500381/2), letter dated 2016 MAR 23 to consignees on 3/28/16. The letter describes the product, problem and actions to be taken. The customers were instructed to review the information with all members of your staff who need to be aware of the contents of this communication. If, at any time, the control panel assembly on your ultrasound system wobbles or feels loose, stop using your system immediately and contact your local Philips representative or Philips Customer Service at 1-800-722-9377. Otherwise, you may continue to use your system. Philips will contact all EPIQ customers to arrange for service to replace the fasteners connecting the control panel assembly to the system with fasteners less likely to loosen with repeated handling. This service will be performed free of charge. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Service at 1-800-722-9377.
Ariol is an automated scanning microscope and image analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the detection, classification, and counting of cells of interest based on particular color, intensity, size, pattern, and shape. This particular Ariol software application is intended to measure, count, and quantitate the percentage and intensity of positively stained nuclei in formalin-fixed paraffin-embedded tissue specimens immunohistochemically sta
An issue during the manufacturing process caused systems to have an improperly activated Windows 7 OS, even though a valid Windows 7 license was associated with each manufactured system.
According to the firm a decision to conduct a Field Corrective Actions of the Ariol Image Analysis and Scanning Systems was done out of an abundance of caution. Leica Biosystems Richmond sent Medical device Field Correction notification letters to the customers of the potentially affected product. The initial written communication was sent to the customers via first class USPS on May 6, 2016. The Medical Device Field Correction letter describes the issues identified with the Ariol Image Analysis and Scanning Systems and the Windows 7 operating system. The letter provided the customer with actions they are required to take which include the following: 1) whether they have a properly activated Windows 7 operating system. 2) If they do not have a properly activated operating system they should to contact Leica Technical Support for additional information and to request a service appointment for correction to their system. For USA customers, they ask that they call 18005374669. International customers are asked to go to http://www.leicabiosystems.com/contactus/contactusonline/ and then choose your country. 3) Customers are asked to confirm receipt of the notice by signing and dating the attached Acknowledgement Form and emailing a copy to RichRA@leicamicrosystems.com or faxing it back to us at 18472363800. Service personnel will be sent to the affected customer locations to replace the hard drives of the affected systems.
CytoVision Image Analysis and Capture System, an automated cell-locating device. The Cytovision is a rapid metaphase finder, image acquisition and computer aided chromosome analysis system which assists the operator in viewing chromosomes and looking for cellular anomalies. CytoVision enables a qualified Cytogeneticist to rapidly and accurately analyze the chromosome banding pattern.
Systems have an improperly activated Windows 7 OS, even though a valid Windows 7 license was associated with each manufactured system. The system appears to be unlicensed. These systems cannot be re-activated using standard Windows licensing procedures and are requiring checks of each potentially affected system to confirm Windows 7 OS is properly activated.
Leica Biosystems Richmond sent Medical device Field Correction notification letters to the customers of the potentially affected product. The initial written communication was sent to the customers via first class USPS on May 2, 2016. According to the firm a decision to conduct a Field Corrective Actions of the Ariol Image Analysis and Scanning Systems was done out of an abundance of caution. The Medical Device Field Correction letter describes the issues identified with the CytoVision Image Analysis and Capture System and the Windows 7 operating system. The letter provided the customer with actions they are required to take which include the following: 1) whether they have a properly activated Windows 7 operating system. 2) If they do not have a properly activated operating system they should to contact Leica Technical Support for additional information and to request a service appointment for correction to their system. For USA customers, they ask that they call 18005374669. International customers are asked to go to http://www.leicabiosystems.com/contactus/contactusonline/ and then choose your country. 3) Customers are asked to confirm receipt of the notice by signing and dating the attached Acknowledgement Form and emailing a copy to RichRA@leicamicrosystems.com or faxing it back to us at 18472363800. Service personnel will be sent to the affected customer locations to replace the hard drives of the affected systems.
Imagecast PACS with Centricity RIS-IC versions prior to 10.6 Update Package 18 Product Usage: is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mamm
A software defect was discovered that causes images to be out of context with clinical information.
GE sent an "Urgent Medical Device Correction" letter to all affected customers on April 2, 2013.The letter identified the product the problem and the action customers need to take. The software has been modified to correct these defects. The modified software will be released and installed in user sites through GE Healthcare FMI #85204. For further questions, please call ( 847) 277-4468.
The GE Centricity Web; Software Version 3.0x, is intended for use under the direct supervision of a licensed healthcare practitioner. The Centricity Enterprise Web is an image and information distribution system for the clinical review of medical images and reports. The Centricity Enterprise Web is not intended for primary diagnosis. The Centricity Enterprise Web is available as a supplemental sub-system to Centricity PACS or as a stand alone Web-based image and information distribution system.
GE Healthcare has become aware of an issue associated with rejected images using the WADO protocol within GE Centricity Web that may impact patient safety. The WADO (Web Access to DICOM Objects) interface allows for embedding DICOM images as JPEGs on a web page and is available as a licensed operation on Centricity Web. When rejected images are archived by Centricity PACS to Enterprise Archive, an
Urgent Medical Device Correction letters dated April 19, 2013 were sent to customers. The letter requests that customers ensure that their system is configured to disable the WADO interface using the included instructions until the software revision can be installed at their site. The letter further informs customers that updated software will be sent to them for installation.The software has been modified to correct these defects. The modified software will be released and installed in user sites through GE Healthcare FMI# 85206R.
Weck Facial Closure Systems have application in laparoscopic procedures for approximation of tissues and percutaneous suturing for closing incision sites.
The devices wings may become partially detached from the EFx Shield during use.
Consignees were notified via letter on 11/16/2016.
MYLA server HP Proliant: DL380-G8 and ML350-G6 version 3.xx computer application software. MYLA V3 is a computer application ("Middleware") based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s).
The MYLA server could slow down due to the volume (weight) of the data to manage and it could also stop processing, impacting the server start-up phase. The problem occurs in high volume settings. The MYLA server slows down and will stop processing completely if any task requires more than 5 minutes to complete. The server then fails to reboot and requires service, provided remotely or on-site
Consignees were notified electronically on April 14, 2015 and letter via FedEx on 04/20/2015.
The Integra Camino ICP Monitor is a compact, portable device for use with Pressure/Temperature, Pressure and Oxygen Partial Pressure Transducer-Tipped Catheters. The CAM02 monitor measures and displays intracranial ICP) and intracranial temperature (ICT).
Integra LifeSciences Corporation is recalling Integra's Camino ICP Monitor (CAM02) & Licox PtO2 Monitor ( LCX02 & LCX02R) due to complaints that the audible speaker malfunctions.
Integra initiated this recall by sending an Urgent Voluntary Medical Device Recall Letter/Recall Acknowledgement Form dated 4/14/15 via traceable courier service to their affected customers.
Edwards Lifesciences IntraClude Intra-Aortic Occlusion Device, model number ICF100, 10.5 Fr (3.5 mm) x 39.4" (100 cm) ThruPort System. Packaged sterile and non-pyrogenic in a sealed, peel-type pouch For use in patients undergoing cardiopulmonary bypass.
Potential device damage on the strain relief near the hub area, which may result in blood leakage.
Urgent Medical Device Safety Notice letters were sent via Fed-Ex by May 05, 2015. As part of the customer instructions in the safety notice, customers were requested to review the notice to understand the potential hazard, complete and return the acknowledgement form, and forward copies of the Safety Notice to any organization where the potentially affected devices have been transferred. The device will only be distributed to customers who have returned the acknowledgement form. Questions should be directed to Edwards Customer Service at 800-424-3278.
TRYX Antibacterial Envelope Rx Only Product Usage: Indicated for stabilization of implanted pacemakers (IPG) and/or implantable cardioverter defibrillators (ICD)
TRYX products are being recalled since the processes of spaying, welding, drying oven and polymer were not adequately validated.
Medtronic sent an Urgent Medical Device Recall letter dated June 2016 to affected customers. The letter identified the affected product problem and actions to be taken. Customers are asked to immediately remove and quarantine all unused product that remains in inventory, return unused product to Medtronic and contacted the local Medtronic representative or Customer Service at 800-848-9300 to assist with the return and credit of unused product. Medtronic will provide credit for all non-expired, unused product. The customer should complete the attached Customer Confirmation Certificate and email to RS.CFQFCA@Medtronic.com or fax to Medtronic at 651-367-0612 Attention Customer Focused Quality.
LightMix Zika rRT-PCR Test The LightMix Zika rRT-PCR Test is a real-time RT-PCR test intended for the qualitative detection of RNA from the Zika virus in serum or EDTA plasma from individuals meeting CDC Zika virus clinical criteria (e.g., clinical signs and symptoms associated with Zika virus infection) and/or CDC Zika virus epidemiological criteria (e.g., history of residence in or travel to a geographic region with active Zika transmission at the time of travel, or other epidemiologic criteri
The LightCycler 480 algorithm used for the LightMix Zika rRT-PCR Test, EUA (catalog number 07987897001) occasionally (~1% of the time) calls a positive result for a negative sample if there is an inflection point in the fluorescence signal.
Roche Molecular Diagnostics Inc. sent an Urgent Medical Device Correction Letter dated November 16, 2016, to all affected customers via UPS. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product until Roche provides updated Instructions for Use to assist in result interpretation. If product was further distributed, ensure the Urgent Medical Device Correction letter is provided to that site. Also, customers were instructed to completed the attached fax form and fax it to 1-888-912-2188. For questions customers should call 1-800-526-1247.
cobas EGFR Mutation Test, v2 and cobas cfDNA Sample Preparation Hungarian Translation Instructions for Use
An error was found within the Hungarian translations of the cobas EGFR Mutation Test v2 Instructions for Use (M/N 07340761001-01HU, Doc Rev. 1.0, Dated 08/2015) and the cobas cfDNA Sample Preparation Kit Instructions for Use (M/N 07573758001-01HU, Doc. Rev. 1.0, Dated 05/2015).
Roche Molecular Diagnostics Inc. issued an Urgent Field Safety Notice dated March 15, 2016 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Actions to be taken by the customer: " If you have utilized 2 L of DNA PBB during testing of plasma specimens, the samples must be retested using the correct volume (2mL DNA PBB). Communication of this Field Safety Corrective Action This notice must be passed on to all those who need to be aware within your organization or to any organization/individual where the potentially affected devices have been distributed/supplied. Please maintain awareness of this notice and resulting action for an appropriate period to ensure the effectiveness of the corrective action. The undersigned confirms that this notice has been notified to the appropriate Regulatory Agency. For questions regarding this recall call 908-253-7200.
cobas EGFR Mutation Test V2 Materials number CE-IVD: EGFR v2: 07248563190 cfDNA: 07247737190 Device Identifier: EGFR v2.0: 00875197005448 cfDNA: 00875197005424 Product Usage: A real time PCR Test for the qualitative detection and identifications of mutations n exons 18, 19, 20 and 21 of the epidermal growth factor receptor (EGFR) gene in DNA derived from formalin-fixed paraffin-embedded (FFPET) tumor tissue or plasma from non-small cell lung cancer (NSCLC) patients. The test also is intended
During internal studies using contrived plasma specimens, several mutations (L8568R, Exon 19 deletion, T790M) inconsistently generated "No Mutation Detected" (ie False Negative ) result with the cobas EGFR mutation Text V2.0 when utilizing the cobas cfDNA Sample Preparation kit.
Roche Molecular Diagnostics sent an Urgent Field Safety Notice letter dated April 5, 2016 to affected customers. Then letter identified the affected product, problem and actions to be taken. The Instructions for use are provided for the user until new ones are available..
Taut Intraducers 10/BX7.5 FR 3.5, Product Code PI-93. To provide an access port into the peritoneal cavity.
The defective component can cause gas leakage through the device. This can result in loss of pneumoperitoneum, potentially necessitating removal of the defective device and insertion of another device.
Recall letters were sent to customers and distributors. The letter requested that the accounts discontinue use and quarantine for return any product. The distributors were requested to conduct a sub-recall. The letters included a Recall Acknowledgement Form which was to be returned to Teleflex.
NOVADAQ SPY-PHI Drape, 7" x 104" (18 cm x 265 cm), Ref. HH2000 (single drape), Sterile, Rx. Manufactured for: Novadaq Technologies Inc., Burnaby, BC, Canada. The SPY-PHI Drape is intended to cover the SPY-PHI System to maintain a sterile field, and as an aid in the clean-up of equipment after surgery or other medical procedures. The product is not intended to be used as patient drapes and they have no patient contact.
The sterile drape material may contain a pinhole compromising its integrity resulting in a breach of the sterile barrier.
The recalling firm, Novadaq, issued an "URGENT DEVICE RECALL" letter dated 6/15/2017 via delivery by firm sales representatives or courier service (such as FedEx or UPS) on 6/27/2017 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to return unused product to Novadaq representative and complete and return the Acknowledgement of Receipt form by email to Novadaq Quality Department at quality@novadaq.com within 24 hours of receipt. Any questions or concerns may also be sent the listed email address. If you have any questions, please do not hesitate to contact the Director of Regulatory Affairs at 1.905.629.3822 ext 205 or email: jpendlebury@novadaq.com.
Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.
The torque wrench used to tighten system bolts during installation was found to be out of tolerance
The recall was initiated via phone and visit beginning 9/18/2017.