FDA Device

The Vantage Digital Panoramic System delivers high-value features standard on every system: - Five pre-programmed exam settings, including bitewing, minimize exam set-up time and allow routine diagnostic exams to be performed extraorally for improved clinical efficiency and patient experience. - Progeny Clarity Enhanced Filter Set provides superior diagnostic imaging quality - Exclusive Focal Trough Alignment reminder assists in consistent patient positioning - Progeny Imaging software included

Hazard

An error was identified in the Real Time Controller (RTC) firmware versions 3.0, 3.1, 3.2, and 3.3 used in Vantage Panoramic X-Ray systems. The error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. failure to stop").

What You Should Do

Remedy

An URGENT MEDICAL DEVICE FIELD CORRECTION letter dated September 20, 2013 was sent to all consignees. Letters sent to distributors include instructions on how to identify the defective units, to quarantine the machine(s), and to contact Midmark at 847-415-9739 or via- e-mail at abavuso@midmark.com. Midmark will either schedule an on-site field correction or arrange for the replacement of the affected device(s) depending on the customer's preference. Midmark further requested the distributors provide Midmark with the names, addresses, phone numbers, and e-mail addresses of any customers that purchased any defective units. Field corrections include the upgrading of RTC firmware to revision 3.4 on the defective units and will be performed as no additional charge by Midmark personnel. Letter sent to practitioners include instructions on how to identify the defective units, instructions on how to temporarily reduce the likelihood of the unintentional movement, and to contact Midmark at 847-415-9739 or via- e-mail at abavuso@midmark.com. Midmark will schedule an on-site field correction by upgrading the firmware to revision 3.4.

Products

The Vantage Digital Panoramic System delivers high-value features standard on every system: - Five pre-programmed exam settings, including bitewing, minimize exam set-up time and allow routine diagnos
Brand
Midmark Corp dba Progeny Inc
UPC
Vantage Panoramic X-Ray System firmware versions: 3.0, 3.1, 3.2, and 3.3.   Vantage Panoramic X-Ray System Catalog Numbers and Serial Numbers: Catalog Number V5000, Serial Numbers ZM81025, ZM81026, ZM81030, ZZ08005, ZZ08006, ZZ08007, ZZ08008, ZZ08009, ZZ08010, ZZ08011, ZZ08012, ZZ08013, ZZ08015, ZZ08016, ZZ08017, Z08018, ZZ08019, ZZ08020, ZZ08021, ZZ08022, ZZ08023, ZB51001, ZB51002, ZB51006, ZB51007, ZT18018, ZY49014, ZK58019, ZS54011, ZP01043, ZV78016, ZE09006, ZN44018, ZS59007, ZS59001, ZW15007, ZW34003, ZW34012, ZW34018, ZY49012, ZY49015, ZZ02010, ZM81022, ZS54012, ZE09019, ZM81017, ZT18004, ZT18015, ZN44020, ZE09012, ZM81023, ZT27012;  Catalog Number V5050, Serial Number ZZ08014;  Catalog Number V5100, Serial Numbers ZM81027, ZM81028, ZZ08003; Catalog Number V5150, Serial Number ZZ08002;  Catalog Number V5150C, Serial Number ZM81016

Units Affected

58 units