Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.
Hazard
The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.
What You Should Do
Remedy
A letter dated 9/17/08 was sent to each account telling them that the PLC would be upgraded. The firm visited each account and made corrections.