Hill-Rom Procedural Stretcher with auto-contour option; Model P8000F. Wheeled hospital stretcher consisting of a platform mounted on a wheeled frame, designed to transport patients and to allow for patient care before, during and after transport.

Hazard

The auto contour function on/off handle may be inadvertently activated while the head of the stretcher is raised, increasing the possibility of the head section to become jammed and preventing it from lowering. This may result in a delay to the procedure.

What You Should Do

Remedy

Consignees will be visited to correct the problem and notified via a hand delivered Device Modification letter dated 4/3/08 at the time of the visit. If you have any questions please contact Jeffrey C. Shaul, Director, RA/QA Acute Care at (812) 931-2604.

Products

Hill-Rom Procedural Stretcher with auto-contour option; Model P8000F. Wheeled hospital stretcher consisting of a platform mounted on a wheeled frame, designed to transport patients and to allow for p
Brand
Hill-Rom, Inc.
UPC
All units with serial numbers H352AN3566 through I222AN9588 and shipped prior to 8/13/07.

Units Affected

564