Roche Diagnostics LDL_C, LDL- Cholesterol Plus 2nd generation, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 03038866322. An in-vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative determination of LDL-Cholesterol concentrations in serum and plasma.
Hazard
Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader.
What You Should Do
Consignees were notified via an Urgent Medical Device correction letter dated 6/11/08 informing users of the problem and providing them with work-around instructions. The instructions suggests 1) inserting the cassette rack slowly; 2) using mainly reagent slots A, B, or C; 3) tilting the rack slightly clockwise so the barcode scanner does not read the barcode at an exact 90 degree angle; and 4) confirming the rack is properly seated on the track and displayed on the Cassette Status screen. If the cassette barcode does not register after following the Work-Around instructions, then customers are required to contact Diagnostics Technical Support.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Roche Diagnostics Corp.
- UPC
- Lot 69740501.
Units Affected
623