PLV-100 Portable Lifecare Ventilator Life support ventilation The device is intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Life support ventilation

Hazard

Certain PLV-100 ventilators may not trigger a signal to activate third-party remote alarms or nurse call systems upon device failure.

What You Should Do

Remedy

The recalling firm issued an URGENT MEDICAL DEVICE RECALL letter dated 7/8/08. The letter informed the customer of the problem and the need to contact the firm for a replacement. Contact Respironics Customer Service at 1-877-387-331 for assistance.

Products

PLV-100 Portable Lifecare Ventilator Life support ventilation The device is intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breat
Brand
Respironics, Inc.
UPC
Model/Catalog number PLV-100. Circuit board part numbers 1034786 version 5 or 1036038 version 3

Units Affected

197