WillowWood Global Llc Recalls

FDA
1
Total Recalls
1
Agencies
May 2021
First Recall
May 2021
Latest Recall

Showing 1-1 of 1 recalls

FDA

Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a one-way air expulsion valve plus a locking mechanism. Product design features an easy release mechanism. Available with or without a titanium pyramid. The lanyard is part of the Alpha Interlock" Suspension System. This system allows you to interchange suspension methods without remaking a socket Model, 3 part numbers - 700-AIS100, Alpha Interlock Suspension System with Alpha L

the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.

On 04/20/2021, Willow Wood issued an Urgent Medical Device Removal notice to customer via letter notifying them it was discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.