TxRx 15Ch Knee Coil 3T QED part number:Q7000008 Siemens Model # 10185460
Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please identify your affected customers and notify them. 3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.
TxRx Knee 15 Coil 3T. For use in conjunction with magnetic resonance scanners to produce diagnostic images of the knee that can be interpreted by a trained physician.
On July 27, 2012, it was reported to the firm that the Siemens 3T 15ch Knee Coil was not functional because the coil identification code for the device was incorrect.
Since all of the devices were located at various Siemens distribution locations, all communications were internal to Siemens concerning bringing the coils from one warehouse to another.