FDA

Atrapro Antipruritic Hydrogel, 1.5 oz gel tube, 4 Net wt/ 1.5 oz.. RX only, Professional Sample, Not for Sale. Pre-printed tube: MS31071, Shelf carton: PL51091-08 Atrapro Antipruritic HydroGel is indicated to manage and relieve the burning, itching and pain experienced with various types of dermatoses, including radiation dermatitis and atopic dermatitis.

Out of specification result - stability test failure of free available chlorine (FAC) which functions as a preservative for the hydrogel. The 13 month test result obtained was 4 ppm FAC and the stability specification is >10 ppm FAC.

Oculus sent an Urgent Medical Device Recall letter dated February 12, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue distribution and segregate the affected product. Immediately remove the affected lots of Recall product from the inventory of sales representative, sub contracted distribution and or fulfillment centers. Segregate the affected product in a secure location for return or destruction. If the product was further distributed customers were instructed to notify healthcare professionals of this recall. Customers were also instructed to completed and return the Business Reply Card and fax to 1-415-462-5163. Once the firm receives the affected product the customer would be contacted by an Oculus Representative and provided a Returns Goods Authorization (RGA) number. For questions customers were instructed to call 1-707-559-2445. For questions regarding this recall 707-559-7191, ext 14.

FDA

Microcyn Solution with Preservatives (Rx), 16.9 fl oz (500 mL) and 33.5 fl oz. (990 mL) Ingredients: Electrolyzed Water (99.97%) Sodium Chloride (NaCl) (230 mg/L) and preservatives: Sodium Hypochlorite (Na0Cl) (40 mg/L), Hypochlorous acid (H0CL) (30 mg/L) -Non Foaming -Does not contain antibiotics - Does not facilitate resistance - No known drug/treatment interactions or contraindications -laboratory proven in solution inactivation of bacteria, viruses, fungi and spores. Performance testing: Mic

Unapproved drug claims and those not approved under 510K clearance.

Oculus Innovative Sciences sent an"URGENT MEDICAL DEVICE RECALL" letter dated November 15, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to identify and remove all affected product from inventory. A Business Reply Card was attached to the letter for customers to complete and return via fax to 1-415-462-5181. Call Oculus Innovative Sciences Customer Service at 1-800-759-9305 for questions regarding this recall.