Del Mar Reynolds Medical, Ltd. Recalls

FDA
3
Total Recalls
1
Agencies
Nov 2015
First Recall
Oct 2017
Latest Recall

Showing 1-3 of 3 recalls

FDA

Arkon Anesthesia Delivery System, Model 99999. Product Usage: Continuous Ventilator The Arkon Anesthesia Delivery System is designed to provide emergency oxygen, vaporized agent delivery and manual ventilation in the event of a power failure scenario. Clinicians are able to use Emergency O2 and manually ventilate the patient, providing gas and agent. However, the anesthesiologist will not have access to mechanical ventilation or ventilator monitoring.

Spacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.

On 08 February 2017, a customer letter will be sent via priority service, return receipt requested, to all U.S. customers. Records of notification will be maintained to ensure complete notification. On 15 February 2017, a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Records of notification will be maintained to ensure complete notification. Upgrade is expected to be available in April 2017; all affected Arkon Anesthesia Delivery Systems will receive a software upgrade onsite.

FDA

Spacelabs Healthcare Arkon Anesthesia Delivery System is intended for use in the hospital environment and operating room. Absorber Adapter Assembly Service Kit, part number: 050-9045-00. Absorber Adapter Assembly, part number: 650-1648-01. Arkon Absorber Canister Carrier, part number: 380-0916-01.

Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.

Spacelabs sent an URGENT - MEDICAL DEVICE CORRECTION letter dated April 13, 2017, to US customers via priority service. The customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record on April 14, 2017. The letter announced that the The Arkon Anesthesia Delivery System, Model 99999, and the Absorber Adapter Assembly Kit part number 050-9045-00, and Absorber Adapter Assembly part number 650-1648-01 are being recalled. Customers are advised that The Arkon Anesthesia Delivery Systems Model 99999 with Absorber Adapter Assembly part number 650-1648-00 is NOT affected by this recall. Spacelabs Healthcare is developing a hardware upgrade to improve the robustness of the affected Absorber Adapter Assembly. A Spacelabs representative will contact consignees once the upgrade is available to schedule a convenient time to update their Arkon Anesthesia Delivery System(s) at no cost. For additional information or technical assistance, please contact: 1-800-522-7025 select 4 for Anesthesia Technical Support

FDA

Spacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcare Blease 700/900 Series Ventilators. Designed specifically for the mechanical ventilation of adult and pediatric patients under general anesthesia.

Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actual delivered volumes being administered to the patient.

Spacelabs started mailing the Urgent Medical Device Correction- Pediatric Flow Sensor kit, PN: 376-0561-00 used with Spacelabs 700 and 900 series ventilators letter, dated 17 March 2015, via priority service to all US customers. Spacelabs will be emailing the customer letter (translated as necessary) to all international subsidiaries and distributors of record on March 20, 2015. Customers are asked to dispose of all pediatric flow sensors purchased from Spacelabs (even if they shipped outside of the time frame in which affected product shipped). Customers are requested to complete and return the Business Reply Form to Spacelabs Healthcare either by fax to Global Tech Support at 425-363-5758 or scan the form and email to techsupport@spacelabs.com Spacelabs will send you at no charge replacement pediatric flow sensors (376-0561-01) for all Spacelabs pediatric flow sensors you purchased. These NEW sensors will be marked with a "-01" molded into the body so you can tell the difference between the old sensors and the new ones. For additional information or technical assistance, please contact: Technical Support Spacelabs Healthcare, Inc. 35301 SE Center Street Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support