Davol Salute Fixation Reusable Handle Item Code: 0113024-Product Code 9113024 (Refurbished)
Malformed constructs (straight shots) may form causing injury to user/patient
Davol issued recall letter notification on 7/16/07 via Federal Express. Sales of the disposable cartridges ceased July 31, 2007. Cartridges will not be requested to be returned to Davol. Customer requests to return cartridges will be managed on an account-by-account basis. A reminder Second Letter issued July 23,2007 as a follow-up to the letter of July 16, 2007. The second stage will be a series of three letters to customers following the cessation of disposable cartridge sales and to recall the re-usable handles. A Stock Status form will be provided to customers to inventory the product which they are returning. The Stock Status form will be used to document the handles being returned. In the event that the customer cannot locate one or more handles. All handles will be returned to Cranston and will be dispositioned. Account verification for the number of handles returned from the account will be performed. No verification testing will be performed on returned handles unless a formal complaint has been registered for a handle.
BARD Simpulse" SOLO System with Soft Splash Shield Tip Product Code: 0067590. The Simpulse SOLO Suction/Irrigator is designed to provide controlled powered irrigation during open orthopedic and general surgical procedures and for the cleansing of burns or wounds. Appropriate fluids for use include fluids with specific gravity less than 1. The pulsatile action of the pump helps to remove blood, tissue debris and foreign matter from the operative or wound site. When connected to a suction source
Product sterility may be compromised due to lack of package integrity.
Davol issued a Recall Letter dated July 30, 2009 requesting user facilities examine inventory and immediately discontinue use of any product from the affected lot and return it to Davol. Direct questions about the recall to Davol, Inc., Sub. C. R. Bard, Inc. by calling 1-401-463-7000.
Davol Salute Fixation System, 38cm Reusable Shaft (Laparoscopic) with Handle. To attach mesh to the abdominal wall. Reference Number 0113036 or 9113036(refurbished)
Inadequate directions for use in the Salute System. Labeling -Revised to provide updated instructions for use and proper servicing
Davol notified users by a first letter on 12/27/06 providng a introductory letter explaining the Enhanced Servicing Program and changes made to the IFU (re: servicing, tip inspection, end of wire, and test fire protocol). A second letter will issue to accounts to identifying the serial number to be serviced and returned to Davol.