Data Innovations, Inc. Recalls
Showing 1-4 of 4 recalls
Data Innovations Instrument Manager Version 8:00, 8.01, 8.02, 8.03 or 8.04 with Specimen Management and using Results/Edit/ReleaseScreen (R/E/R)
Patient results may be associated with an incorrect specimen
Data Innovations notified customers by an email listserv notification and by letter dated 8/17/06. A patch would be available on the website to be downloaded.
Roche Modular Drive (rchmdlri), analytical laboratory data interface.
Software of modular driver may incorrectly report patient results as Quality Control Results
Data Innovations notified users by email listserv and by letter dated 10/25/06.
Instrument Manager driver : Driver sysu100i that connects to the Sysmex UF-100 analyzer Software all versions prior to( v7.00.0006 and v8.00.0005): v7.00.0001, v7.00.0002, v7.00.0003, v7.00.0004, v7.00.0005, v8 .00.0001, v8.00.0002, v8.00.0003, and v8 .00.0004
When connected to the Sysmex UF-100, the Instrument Manager driver sysu100i driver software versions 7.00.0001, 7.00.0002, 7.00.0003, 7.00.0004, 7.00.0005, 8.00.0001, 8.00.0002, 8.00.0003, and 8.00.0004, and is configured with option "Strip Zeros from Result" set to "Yes" may result in sample error
Data Innovation notified consignees by Important Medical Device Recall Notification dated 11/5/13 to all list serv recipients of the problem and posted critical alert on Customer Web Portal. A certified letter was issued to impacted customers and consignees if no response was recieved from e-mail.
Instrument Manager -Software Versions 8.05, 8.06, 8.07, 8.08, and 8.09. It is a stand-alone software package that provides connectivity between laboratory instruments and Laboratory Information Systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information. Instrument Manager also provides data management functions that aid in the interpretation and management of laboratory results. The Instrument Manager software is put on
The "previous result" retrieved from the Specimen Management Database may not be the most recent "previous result".
Data Innovations sent an "IMPORTANT PRODUCT NOTICE" dated May 11, 2009 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter offers temporary workarounds for customers to use until corrected versions of the software is complete. Versions of the corrections will be provided on a staggered schedule. Customers will be notified by e-mail. Questions regarding this notice are directed to Support at 802-658-1955.