Abaxis, Incorporated Recalls

FDA
1
Total Recalls
1
Agencies
Aug 2011
First Recall
Aug 2011
Latest Recall

Showing 1-1 of 1 recalls

FDA

Abaxis brand Piccolo Comprehensive Metabolic Panel; Used with the Piccolo Blood Chemistry Analyzer or Piccolo xpress Chemistry Analyzer; Catalog/Part Number: 400-0028; Product is manufactured and distributed by Abaxis, Incorporated, Union City, CA The Piccolo Comprehensive Metabolic Reagent Disc, used with the Piccolo Blood Chemistry Analyzer or Piccolo xpress Chemistry Analyzer, is intended to be used for the in vitro quantitative determination of alanine aminotransferase (ALT), albumin, al

Abaxis has determined that during the manufacture of product lot 0396BC5 may report inaccurate values of Alanine Aminotransferase (ALT) when used in the Piccolo Blood Chemistry Analyzer or Piccolo xpress Chemistry Analyzer.

The firm, Abaxis Inc., sent an "URGENT PRODUCT RECALL NOTICE" dated November 19, 2010, to all customers. The notice described the product, problem and action to be taken by the customers. The customers were instructed to check their inventory of the lot for boxes #0665 or higher; if found, to discontinue use; contact Abaxis for immediate replacement; complete and return the enclosed PRODUCT RECALL RESPONSE form to Abaxis Technical Support via fax at 1-877-349-2087, and ensure that their laboratory staff is informed of this recall notice. If you have any questions regarding this Product Recall Notice, please contact Abaxis Medical Technical Support at 800-822-2947, or at medtechsupport@abaxis.com.