TCL1 Breakapart Probe, Fluorescence in situ hybridisation (FISH) is a technique allows the visualisation of DNA sequences upon chromosome Catalog Number: LPH046-A
Hazard
May show unexpected locus specific signals in addition to those at 14q32.
What You Should Do
Cytocell issued Urgent Medical Device Recall letter (VC/2022/005) to Distributor/End User via email on 2/4/22 . Letter states reason for recall, health risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that you destroy the remaining inventory. We also suggest that laboratories undertake a review of the results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of any additional locus specific signals at 9q34. Please complete and return the enclosed response form as soon as possible. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Cytocell Ltd. 418 Cambridge Science Park, Milton Road Oxford Gene Technology Cambridge United Kingdom
- UPC
- Lot 074612
Units Affected
1 unit