FDA Device

Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities

Hazard

When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.

What You Should Do

Remedy

Sensus healthcare issued an "Urgent Device Correction" to its consignees on 02/10/2023 by email. The notice requested the following actions: "Do not initiate beam delivery after a pause & resume sequence without verifying the amount of dose delivered. Do evaluate any patient treatments with version 1.9.2 that followed the pause & resume workflow. Do verify that the amount of dose to be delivered is as expected after each pause & resume sequence." The firm has developed a permanent software resolution for this issue. Customer service will contact the customer to coordinate upgrading to this revised software. Questions: Contact Sensus Customer Support at support@sensushealthcare.com or 561-578-4982

Products

Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities
Brand
Sensus Healthcare, Inc.
UPC
Software version 1.9.2 loaded onto Serial Numbers: 1512-2002; 1512-2003; 1606-2007; 1609-2010; 1609-2011; 1609-2018; 1612-2023; 1612-2025; 1705-2032; 1709-2046; 1803-2068; 1906-2137; 1909-2149; 1912-2167; 2003-2177; 2112-2201; 2112-2202; 2112-2208; 2112-2218; 2203-2232; 2203-2241; 2206-2264; 2206-2272

Units Affected

23 units