CK MB Calibrator, Catalogue Number CK2393
Hazard
Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.
What You Should Do
Remedy
Firm notified customers via email on March 25, 2022. The distributors and customers were instructed to discard previous target sheet and perform calibration with the realigned target value provided with the letter. Also, to complete and return the response form 12187-QA to technical.service@randox.com.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
CK MB Calibrator, Catalogue Number CK2393
- Brand
- Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
- UPC
- GTIN 05055273201451 Lot 4410CK
Units Affected
1,392