FDA Device

JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035821 SIZE 2, CATALOG NUMBER: REF 74035822 SIZE 3, CATALOG NUMBER: REF 74035823 SIZE 4, CATALOG NUMBER: REF 74035824 SIZE 5,

Hazard

The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.

What You Should Do

Remedy

The firm initiated the recall by email on 01/28/2020. The initial consignee notified the district sales representatives on 01/28/2020 by email. The district sales representatives checked their inventory and the inventory of the hospital they service.

Products

JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to asses
Brand
Smith & Nephew, Inc.
UPC
SIZE 1, CATALOG NUMBER: REF 74035821; LOT/BATCH CODES: 19EM19112R, 19FM05587R, 19FM11174R, 19HM18075, 19HM19648, 19HM21015, 19HM21016, 19HM23242  SIZE 2, CATALOG NUMBER: REF 74035822; LOT/BATCH CODES: 19EM21573R, 19EM21573S, 19FM05589R, 19HM21031, 19HM21033, 19HM21878, 19HM21881  SIZE 3, CATALOG NUMBER: REF 74035823; LOT/BATCH CODES: 19EM21578R, 19HM19009, 19JM09476, 19JM18509, 19JM18525, 19JM21072, 19KM14473, 19KM17780  SIZE 4, CATALOG NUMBER: REF 74035824; LOT/BATCH CODES: 19EM14365R, 19EM14365S, 19FM11173, 19HM18078, 19HM19659, 19HM21861, 19JM21075, 19KM15573, 19KM1557  SIZE 5, CATALOG NUMBER: REF 74035825; LOT/BATCH CODES: 19EM14367R, 19HM18079, 19HM21855, 19KM12683   SIZE 6, CATALOG NUMBER: REF 74035826; LOT/BATCH CODES: 19EM24480R, 19HM18080, 19HM21856, 19KM21512  SIZE 7, CATALOG NUMBER: REF 74035827; LOT/BATCH CODES: 19EM14371R, 19FM01063R, 19FM11177R, 19HM18335, 19HM21873, 19HM21874, 19KM17781  SIZE 8, CATALOG NUMBER: REF 74035828; LOT/BATCH CODES: 19EM24498R, 19EM24498S, 19FM05591R, 19FM11178R, 19FM11178S, 19JM07464, 19JM07465, 19KM07278, 19KM12699  SIZE 9, CATALOG NUMBER: REF 74035829; LOT/BATCH CODES: 19EM21582R, 19FM11179R, 19GM21787, 19JM07467, 19KM07276, 19KM15575, 19KM15576  SIZE 10, CATALOG NUMBER: REF 74035830; LOT/BATCH CODES: 19EM14372R, 19FM11180, 19HM18254, 19JM07468, 19JM07470, 19JM07472, 19KM05610, 19KM15571

Units Affected

200 units