Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.
Hazard
Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
What You Should Do
Remedy
The firm initiated their correction on 12/16/2019 by letter. The letter explained the issue and directed the consignee to follow instructions in the letter until an update to the Centricity Universal Viewer can be made by the firm to correct the issue.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.
- Brand
- GE Healthcare, LLC
- UPC
- Model 2088026-9XX Versions 6.0 SP10 or higher
Units Affected
205