FDA Device

AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: positioning fluoroscopy procedures, gastro-intestinal examinations, urogenital tract examinations, and angiography. It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system

Hazard

Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible

What You Should Do

Remedy

Agfa issued Urgent Field Safety Notice via email on 4/21/20 to consignees, letter states reason for recall, health risk and action to take: Acknowledgment, via a feedback form or email, that the information was received and understood was requested from the consignees.

Products

AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoros
Brand
Agfa N.V. Septestraat 27 Mortsel Belgium
UPC
Serial Numbers: 0925, 0945, 0948, 0949, 1003, 1022, 1116, 1136, 1161, 1229, 1239, 1401, and 1404

Units Affected

13 units