AGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Usage: TomoSynthesis is used to synthesize tomographic slices from a single tomographic sweep. The DR 800 is not intended for mammography applications
Hazard
Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.
What You Should Do
AFGA issued via email On February 10, 2020, an URGENT FIELD SAFETY NOTICE letter. Letter states reason for recall, health risk, and action to take: With this letter, Agfa is informing you that the DR 800 system still can be used including for Tomosynthesis for mAs values which are larger than 0,2 mAs. In case Tomosynthesis image acquisition sequence does not stop automatically after the expected number of exposures, release the exposure button as this will stop the sequence and notify your Agfa Service contact at once. Contact your local Agfa Imaging organization 1-877-777-2432 and please reference Problem Record PRB2000387 and Vigilance Record VR0000716. Complete/Return by fax or email the URGENT FIELD SAFETY NOTICE Feedback form
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Agfa N.V. Septestraat 27 Mortsel Belgium
- UPC
- Serial Numbers: 00949, 1161
Units Affected
2 units