FDA Device

Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.

Hazard

There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.

What You Should Do

Remedy

The recalling firm issued an Advisory Notice Information dated 2020-06-30 on 7/2/2021 via email which explained the reason for recall and that they would be contacted as soon as possible to proceed with the exchange. On 9/24/2021, a Returned Goods Authorization Form was sent to the consignees so the adapter could be exchanged.

Products

Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.
Brand
Laboratoires Bodycad, Inc.
UPC
Torque Limiting Adaptor - Lot number SB000430.  Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.

Units Affected

10 in the U.S.