FDA Device

BodyGuardian Heart Remote Monitoring Kit, consisting of two BG Heart monitors, either Models BC-1001-A or BC1001-D, and a cell phone, either Models BP1008 or BP10014, that is paired to both of the monitors.

Hazard

The device data being collected and transferred to the monitoring center may not be accurate due to nonvalidated association between the phone software and the heart monitors, therefore, the patient's report should not be used to evaluate their condition.

What You Should Do

Remedy

On 8/6/2019, the recalling firm contacted or attempted to contact by phone all patients that had potentially impacted devices. Phone communication to the physicians of the patients also began on 8/6/2019. All patients that agreed were sent new devices to replace the affected one to finish their study.

Products

BodyGuardian Heart Remote Monitoring Kit, consisting of two BG Heart monitors, either Models BC-1001-A or BC1001-D, and a cell phone, either Models BP1008 or BP10014, that is paired to both of the mon
Brand
Preventice Services, LLC
UPC
Version 2.3 phone software, kit IDs BG2570727, BG2565845, BG2567817, BG2566832, BG2566830, BG2570699, BG2566530, and BG2570700.

Units Affected

8 kits