Monaco TRP System The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Hazard

Contours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm.

What You Should Do

Remedy

Elekta sent an Important Field Safety Notice 382-01-MON-007 was sent to all affected customers on November 17, 2016, to all affected customers. The notice informs users of the specific product version numbers affected by the issue, and any work arounds that can be used to avoid the issue. The notice included a reply form which is to be returned.

Products

Monaco TRP System The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.
Brand
Elekta, Inc.
UPC
Software Builds 5.00 and higher

Units Affected

2148