Monaco TRP System The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.
Hazard
Contours are not handled correctly if the CT dataset contains slices with DICOM Z coordinates of xx.x50mm.
What You Should Do
Remedy
Elekta sent an Important Field Safety Notice 382-01-MON-007 was sent to all affected customers on November 17, 2016, to all affected customers. The notice informs users of the specific product version numbers affected by the issue, and any work arounds that can be used to avoid the issue. The notice included a reply form which is to be returned.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Monaco TRP System
The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy.
- Brand
- Elekta, Inc.
- UPC
- Software Builds 5.00 and higher
Units Affected
2148