FDA Device

Elastic Traction Device, 11mm Lower. UPN: ER56041 Reference Number: ETD-D-26-235-C-N - Product Usage: indicated for use in flexible Endoscopy to provide retraction to assist in tissue resection, exposure, and removal of tissue within the gastrointestinal tract.

Hazard

FDA has informed Micro-Tech that the Elastic Traction Device is not appropriately classified as a Class I device and requires 510(k) clearance to be marketed in the US.

What You Should Do

Remedy

On October 7, 2020, the firm distributed Urgent Medical Device Recall letters to affected customers, advising them that the firm is voluntarily recalling the Elastic Traction Device. Further distribution or use of any remaining product should cease immediately. Customers were asked to identify and quarantine devices left in inventory. They should complete the customer Acknowledgement and Receipt Form accompanying the letter and return the form to Micro-Tech Endoscopy USA. If product is available for return, the firm will contact your facility to create a Returned Material Authorization with shipping instructions. Customers should pass the notice on to all those who need to be aware within their organization, or to any organization where the affected devices have been transferred.

Products

Elastic Traction Device, 11mm Lower. UPN: ER56041 Reference Number: ETD-D-26-235-C-N - Product Usage: indicated for use in flexible Endoscopy to provide retraction to assist in tissue resection, expos
Brand
Micro-Tech Usa
UPC
Lots M190808231 M191018231 M191224231 M200305232

Units Affected

2420 pieces