FDA Device

NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115

Hazard

Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples

What You Should Do

Remedy

Randox notified the distribution center within the USA via email on 5/10/21.The distributor will then contact the customer directly. Letter states reason for recall, health risk, and action to take: Action to be taken: " Please discard previous value sheets and ensure standard value is updated. " Review results generated with the affected batches in line with the clinical profile of the patient. " Discuss the contents of this notice with your Medical Director

Products

NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115
Brand
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
UPC
Batch Number: 544642 Exp. Date: 28th Nov 2022 GTIN: 05055273203066

Units Affected

71 kits US