VITROS Chemistry Products TP Slides- IVD measure total protein (TP) concentration in serum and plasma using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems Product Code: 8392292
Hazard
Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides
What You Should Do
Ortho-Clinical Diagnostics (Ortho) issued customer letter (Ref. CL2018-308) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) on 15 Dec 2020 to all customers who received one of the affected products in the previous 18 months. Letter states reason for recall, health risk and action to take: Complete the enclosed Confirmation of Receipt form no later than insert date. Please provide the enclosed communication to customers who were shipped the affected products (listed above) from your facility within the last 18 months. Ortho has communicated with your customers that received drop shop product from our warehouse. Foreign affiliates were informed by email on 15 Dec 2020 of the issue and instructed to notify their consignees of the issue and their required action until the updated the IFU is available.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Ortho Clinical Diagnostics Inc
- UPC
- All Current, Future and Past Lots UDI: 10758750004850
Units Affected
US= 223,788 ; OUS:=168,120 Totlal: 491,908